Key facts
Start here

Agilent 6530 Accurate-Mass Q-TOF LC/MS System with Agilent 1290 Infinity LC for Multi Plant Growth Regulator Analysis from Grapes handbook
This document presents a sensitive and effective method for screening 12 Plant Growth Regulators (PGRs) in grape samples using the Agilent 6530 Accurate-Mass Q-TOF LC/MS system coupled with the Agilent 1290 Infinity UHPLC. The method utilizes a simple, inexpensive dispersive C18 sample cleanup technique to remove background matrix, enabling the detection of multiclass PGRs with outstanding mass accuracy (less than 1 ppm for 50 ppb and 100 ppb concentrations).
The analysis involves liquid phase extraction followed by dispersive C18 cleanup, internal standard spiking, UHPLC gradient separation, and detection via Jetstream ESI-LC/MS/MS. The document details specific chromatographic conditions, mass spectrometer settings, and results for both positive and negative ion modes, demonstrating the system's capability to identify compounds using accurate mass and structural elucidation via MS/MS.
Agilent 6530 Accurate-Mass Q-TOF LC specifications
| Total run time (positive mode) | 16 min |
|---|---|
| Total run time (negative mode) | 6 min |
| Injection volume | 2 yL |
| Column temperature | 40 °C |
| Flow rate | 0.2 mL/min |
| Mass accuracy (50 ppb/100 ppb) | less than 1 ppm |
| Mass accuracy (5 ppb) | less than 5 ppm |
All Agilent 6530 Accurate-Mass Q-TOF LC documents by type
How to use the product for one specific task, with worked examples.
Questions answered in these manuals
What PGRs are detected in positive mode?
The PGRs detected in positive mode are Chlormequat, d4 chlormequat, Daminozide, Zeatin, Kinetin, 6-Benzyl adenine, Indol Butyric acid, Forchlorfenuron, and Paclobutrazol.
What PGRs are detected in negative mode?
The PGRs detected in negative mode are Gibbrelic acid, Abscisic acid, and 2, 4 Dichlorphenoxy acetic acid.
What is the limit of quantitation (LOQ) for cystine in WBCs?
The LOQ of cystine in WBCs is 0.02 uM, or 40 fmol on-column.
From: Agilent 6530 Accurate-Mass Q-TOF LC/MS System Application Note
What is the dynamic range of the cystine calibration curve?
The cystine calibration curve shows excellent linearity in the dynamic range of 0.02 - 4 uM.
From: Agilent 6530 Accurate-Mass Q-TOF LC/MS System Application Note