Key facts
- Brand
- Agilent
- Model
- 5975 Series Mass Selective Detectors
- Documents
- 4
- Languages
- EN
- Last added
- 2012-06-08
Start here
Agilent 5975 Series Mass Selective Detectors Hardware Installation (Japanese)
This document provides installation precautions and safety tips for the Agilent Technologies 5975 Series Mass Selective Detector hardware. The document includes general information, safety precautions, installation steps, and checklists.
Read onlineAgilent 5975 Series Mass Selective Detectors specifications
| Model Number | G3170A, G3171A, G3172A, G3174A |
|---|---|
| Voltage Fluctuation Tolerance | +5% or -10% from rated voltage |
| Environmental Category | Category II |
Source: Agilent 5975 Series Mass Selective Detectors Site Preparation Guide
All Agilent 5975 Series Mass Selective Detectors documents by type
Mounting, wiring and first power-up. Read it before the hardware goes in.
Questions answered in these manuals
Who is authorized to perform the installation?
Installation must be performed by an Agilent Technologies service representative.
From: Agilent 5975 Series Mass Selective Detectors Hardware Installation Manual
What type of pump does the MSD use?
The 5975 Series Mass Selective Detectors are turbo pump MSDs.
From: Agilent 5975 Series Mass Selective Detectors Hardware Installation Manual
How is the 5975 Series Mass Selective Detector identified?
Each MSD is identified by a unique 10-character serial number located on a label on the lower left side near the front of the instrument.
From: Agilent 5975 Series Mass Selective Detectors Site Preparation Guide
What safety precautions are needed for handling MSD covers?
Never remove a cover unless the detector, inlet, or oven are turned off to avoid contact with dangerous voltages.
From: Agilent 5975 Series Mass Selective Detectors Site Preparation Guide
What safety precautions should I take when handling the 5975 MSD?
Many internal parts carry dangerous voltages; never remove covers unless instructed. Wear a grounded wrist strap to prevent electrostatic discharge. Avoid touching hot parts and use gloves and wrenches when necessary.
From: Agilent 5975 Series Mass Selective Detectors Site Preparation Guide
What are the electrical requirements for installing the 5975 MSD?
The instrument requires a stable power supply with voltage fluctuations no more than +5% or -10% from rated voltage. It must be installed in a Category II environment as defined in IEC 664.
From: Agilent 5975 Series Mass Selective Detectors Site Preparation Guide