ManualLib has 3 documents for the Agilent 5973N MSD: 2 application notes, 1 other. Every file can be read online or downloaded as a PDF for free.

Application Note 2Other 1

Key facts

Brand
Agilent
Model
5973N MSD
Documents
3
Languages
EN
Last added
2012-07-05

Start here

Agilent 5973N MSD Ammonia Reagent Gas Implementation Guide

Agilent 5973N MSD Ammonia Reagent Gas Implementation Guide

Application Note 4 pages · 1.8 MB · EN

This document details the implementation of ammonia reagent gas for positive and negative chemical ionization on the Agilent 5973N Mass Spectrometry Detector (MSD). It highlights that ammonia is useful for positive chemical ionization due to its high proton affinity and for electron capture negative ion chemical ionization due to its thermalizing power. The text emphasizes the necessity of using high-purity ammonia (99.998% or greater) to prevent issues like split peaks and short filament lifetimes caused by water and oxygen contaminants.

Read online

Agilent 5973N MSD specifications

Ammonia Purity Requirement99.998% or greater
Water and Oxygen Limitno more than 5 ppm
Ammonia Usage Rateapproximately 1 ml/min
Regulator Pressure Settingless than 20 psi
Pump Ballasting Duration30 minutes
Pump Oil Change Frequencyat least every three months

Source: Agilent 5973N MSD Ammonia Reagent Gas Implementation Guide

All Agilent 5973N MSD documents by type

Application Note 2

How to use the product for one specific task, with worked examples.

Questions answered in these manuals

What is the minimum purity required for ammonia reagent gas?

The ammonia must be of the highest purity available, certainly no less than 99.998% pure.

From: Agilent 5973N MSD Ammonia Reagent Gas Implementation Guide

What are the critical contaminants to avoid in ammonia gas?

Water and oxygen are the most common and critical contaminants, which should be present at no more than 5 ppm.

From: Agilent 5973N MSD Ammonia Reagent Gas Implementation Guide

Which reagent gas is recommended for reducing chemical noise in phthalate analysis?

Ammonia is recommended as the reagent gas to reduce chemical noise and enhance identification.

From: Agilent A New Approach to the Analysis of Phthalate Esters by GC/MS Application Note

What is the primary benefit of using retention-time locking (RTL)?

RTL allows for the creation of compound acquisition methods and quantitation databases that can be reproduced in any laboratory anywhere.

From: Agilent A New Approach to the Analysis of Phthalate Esters by GC/MS Application Note

What is the purpose of this manual?

This book is used to learn how to operate and maintain your 5973N MSD hardware.

From: Agilent 5973N MSD and 6890 GC Quick Reference Manual

How do I access the online help?

Select Help Topics from the Help menu in any window or click the Help button on any dialog box.

From: Agilent 5973N MSD and 6890 GC Quick Reference Manual