Agilent P200 ScreenTape Suppression of Antibody Reduction Artifacts Technical Overview

Agilent P200 ScreenTape Suppression of Antibody Reduction Artifacts Technical Overview

Product Leaflet 2 pages · 696 KB · EN

This document provides a technical overview regarding the analysis of monoclonal antibodies (mAbs). It addresses the issue where certain IgG subtypes exhibit higher than expected half-antibody representation under denaturing, non-reducing conditions due to cryptic cysteins exposed by dodecyl sulfate.

The text details an experiment using N-ethylmaleimide (NEM) prior to SDS-PAGE analysis. The results show that treating labile IgGs with NEM reduces the formation of these denaturing artifacts on P200 ScreenTape under non-reducing conditions, offering a more accurate method for mAb quantification compared to traditional methods.

Read online

Agilent P200 specifications

Staining reaction optimal pH rangepH 8-10
P200 pH buffer pHpH 9
Urea compatibility limit8M
Guanidine compatibility limit600 mM
Salt concentration limit for migration500 mM
Triton X-100 compatibility limit1%
Acetonitrile compatibility limit20%
Methanol compatibility limit20%
Ethanol compatibility limit20%

Source: Agilent P200 ScreenTape Technical Overview

All Agilent P200 documents by type

Quick Start Guide 1

The shortest path to a working setup, usually only a few pages.

Questions answered in these manuals

How does NEM treatment affect antibody analysis on P200 ScreenTape?

Treatment with N-ethylmaleimide (NEM) prior to SDS-PAGE analysis reduces the appearance of half antibodies by alkylating free cystein residues, preventing auto-reduction.

From: Agilent P200 ScreenTape Suppression of Antibody Reduction Artifacts Technical Overview

What is the recommended pH for the NEM reaction?

The reactions should be carried out at a neutral pH because elevated pH can cause NEM to react with lysine residues and produce unexpected results.

From: Agilent P200 ScreenTape Suppression of Antibody Reduction Artifacts Technical Overview

What is the optimal pH range for the staining reaction?

The staining reaction occurs optimally between pH 8-10 as the amine group needs to be deprotonated to be available to the dye.

From: Agilent P200 ScreenTape Technical Overview

Can guanidine be used in sample preparation?

Guanidine has low compatibility at 600 mM because it causes SDS precipitation. It is recommended to dilute guanidine to below 600 mM or use urea instead.

From: Agilent P200 ScreenTape Technical Overview

Can P200 ScreenTape analyze samples under different conditions?

Yes, it allows analysis of samples under both reducing and non-reducing conditions.

From: Agilent P200 ScreenTape Application Note for Antibody Testing

How many samples can be analyzed without manual intervention?

Integrity and purity analysis of up to 16 mAb preparations can be performed in less than one minute per sample.

From: Agilent P200 ScreenTape Application Note for Antibody Testing