Agilent P200 ScreenTape Technical Overview

This document introduces the effects of proteins and buffers on staining, denaturation, and electrophoresis when analyzing proteins with Agilent P200 ScreenTape


Agilent P200 ScreenTape Technical Overview - cover page
Brand
Agilent
Category
Other
Document type
Other
Language
English
Pages
4
File format
PDF
File size
567 KB
Published
05 July, 2012
Updated
30 September, 2023
Model
P200
MD5 checksum
FD61D74D42E53EBFEAF954A29171B422

About this manual

This document serves as a technical overview for analyzing proteins using the Agilent P200 ScreenTape and 2200 TapeStation system. It details how various protein properties, such as amino acid composition and modifications, can impact the staining reaction and overall analysis results. The text also highlights the variability introduced by different buffer compositions and additives used in protein preparation.

What's inside

  • Introduction
  • Sample handling
  • Impact of protein properties on staining reaction
  • Influence of pH on staining
  • Denaturation
  • Sample conductivity affects migration
  • Buffer compatibility table

Specifications

Staining reaction optimal pH rangepH 8-10
P200 pH buffer pHpH 9
Urea compatibility limit8M
Guanidine compatibility limit600 mM
Salt concentration limit for migration500 mM
Triton X-100 compatibility limit1%
Acetonitrile compatibility limit20%
Methanol compatibility limit20%
Ethanol compatibility limit20%

Frequently asked questions

What is the optimal pH range for the staining reaction?

The staining reaction occurs optimally between pH 8-10 as the amine group needs to be deprotonated to be available to the dye.

Can guanidine be used in sample preparation?

Guanidine has low compatibility at 600 mM because it causes SDS precipitation. It is recommended to dilute guanidine to below 600 mM or use urea instead.

How does high salt concentration affect electrophoresis?

High salt levels retard the dye front, which can affect sizing accuracy. Levels above 500 mM begin to retard the internal 10 kDa marker.

What should be done if a sample has high viscosity due to sugars?

Care must be taken when pipetting small volumes of high viscosity samples to ensure reproducibility of analysis.