Agilent 920-LC Manuals

ManualLib has 4 documents for the Agilent 920-LC: 2 user manuals, 1 programming manual, 1 datasheet. Every file can be read online or downloaded as a PDF for free.

User Manual 2Programming Manual 1Datasheet 1

Key facts

Brand
Agilent
Model
920-LC
Industry
Industrial
Documents
4
Languages
EN
Last added
2012-07-04

Start here

Varian 920-LC Analytical Liquid Chromatograph Specification Sheet

Varian 920-LC Analytical Liquid Chromatograph Specification Sheet

Product Leaflet 4 pages · 130 KB · EN

The Varian 920-LC is an integrated, pre-configured analytical high performance liquid chromatograph (HPLC) system. It features a low pressure quaternary gradient reciprocating dual piston pump with a built-in degasser and an autosampler with direct injection. The system can be controlled by Varian Galaxie Chromatography Software on a single computer or through a networked client and acquisition server.

Read online

Agilent 920-LC specifications

Primary piston displacement100 ul
Maximum pressure (0.001 to 5 mL/min)39.2 MPa (5685 psi, 392 bar)
Maximum pressure (5 to 9.999 mL/min)20 MPa (2900 psi, 200 bar)
Flow rate range0.001 to 9.999 mL/min
Wavelength range (UV-Vis)190 to 900 nm
Wavelength range (PDA)190 to 950 nm
Excitation wavelength range (Fluorescence)200 to 850 nm
Emission wavelength range (Fluorescence)250 to 900 nm
Settable temperature range (Autosampler)4 to 35 °C

Source: Varian 920-LC Analytical Liquid Chromatograph Specification Sheet

All Agilent 920-LC documents by type

Questions answered in these manuals

What is the maximum number of solvents supported?

The system supports a maximum of 4 solvents.

From: Varian 920-LC Analytical Liquid Chromatograph Specification Sheet

What is the injection repeatability of the autosampler?

The injection repeatability is less than 0.3% RSD using a 10 ul injection.

From: Varian 920-LC Analytical Liquid Chromatograph Specification Sheet

What is the run time for the fast HPLC method described?

The run time can be reduced to below 5 minutes using the shorter column.

From: Agilent Productivity Enhancement of USP Assay for Ibuprofen Oral Suspension with the 920-LC Pursuit XRs C8 Column

Does the fast method meet USP requirements?

Yes, the performance surpasses the USP method requirements for resolution, tailing, and precision.

From: Agilent Productivity Enhancement of USP Assay for Ibuprofen Oral Suspension with the 920-LC Pursuit XRs C8 Column

What is the limit for ibuprofen-related compound C in USP monographs?

The amount of ibuprofen-related compound C must be not more than 0.1% per tablet.

From: Agilent Efficient HPLC Analysis in the USP Assay of Ibuprofen in the Limit Test of Ibuprofen-related Compound C in Ibuprofen Tablets with the 920-LC Pursuit C18 Column

How does the optimized method improve productivity compared to the original USP method?

The run times can be reduced to below 4 minutes, which is less than half the run times from the original method.

From: Agilent Efficient HPLC Analysis in the USP Assay of Ibuprofen in the Limit Test of Ibuprofen-related Compound C in Ibuprofen Tablets with the 920-LC Pursuit C18 Column