Agilent Productivity Enhancement of USP Assay for Ibuprofen Oral Suspension with the 920-LC Pursuit XRs C8 Column

This article presents a method for the USP assay of ibuprofen oral suspension using Varian 920-LC and Pursuit™ XRs C8 column. The results obtained from the experiment show that all performance requirements of the USP methods are satisfied and demonstrate a significant enhancement in laboratory productivity when using this system.


Agilent Productivity Enhancement of USP Assay for Ibuprofen Oral Suspension with the 920-LC Pursuit XRs C8 Column - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
3
File format
PDF
File size
107 KB
Published
12 June, 2012
Updated
29 September, 2023
Model
920-LC
MD5 checksum
7C6DEAEF150CCE90FB137A8C30BEDD33

About this manual

This document describes a productivity-enhanced High-Performance Liquid Chromatography (HPLC) method for the assay of ibuprofen in oral suspension, complying with United States Pharmacopoeia (USP) standards. The method utilizes the Varian 920-LC liquid chromatograph equipped with a diode array detector and an autosampler. The study compares a standard method using a 150x4.6 mm column with a faster method using a shorter 50x4.6 mm column to improve sample throughput.

What's inside

  • Introduction
  • Instrumentation
  • Materials and Reagents
  • Standard and Sample Preparation
  • Chromatographic Conditions
  • Results and Discussion
  • Conclusion

Specifications

Column (Fast Method)Pursuit XRs C8, 3 um, 50x4.6 mm
Column (Standard Method)Pursuit XRs C8, 5 um, 150x4.6 mm
Mobile Phase0.01 M Phosphoric Acid: Acetonitrile (55:45 % v/v)
Flow Rate2 mL/min
Wavelength220 nm

Frequently asked questions

What is the run time for the fast HPLC method described?

The run time can be reduced to below 5 minutes using the shorter column.

Does the fast method meet USP requirements?

Yes, the performance surpasses the USP method requirements for resolution, tailing, and precision.

What instrument was used for the experiment?

The experiment used a Varian 920-LC liquid chromatograph with a low pressure quaternary gradient pump.