Agilent 7890A GC Safety Manual
The Agilent 7890A GC Safety Manual provides essential safety information for operating and maintaining the gas chromatograph. It details important safety warnings, high‑voltage hazards, and required personal protective equipment.
The manual also includes specific sections on hydrogen safety, micro‑electron capture detector precautions, fuse and battery specifications, regulatory compliance, cleaning procedures, and product recycling guidance. This manual is written in Chinese.
Agilent 7890A specifications
| Fuse rating (F1, F2) | 20 A, 250 Vac, IEC 127 type f |
|---|---|
| Fuse rating (F3, F4) | 8 A, 250 Vac, IEC type (no time delay) |
| Logic board battery | 3 V Li‑polymer carbon cell, 0.048 A‑h, Panasonic BR‑1225 |
| Sound pressure level | Lp < 70 dB(A) (DIN‑EN 27779) |
| EMC compliance | Meets CISPR 11 requirements |
Source: Agilent 7890A GC Safety Manual
All Agilent 7890A documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
- Agilent 7890A/5975A GC/MSIn the system Automatic Backflush Manual
- Agilent Italian Version - Agilent 7890A Network Gas Chromatograph Data Sheet Manual
- Agilent Italian Version: TheAgilent 7890A Gas Chromatograph Brochure Manual
- Agilent German Version: The Agilent 7890A Gas Chromatograph Brochure Manual
- Agilent French Version: The Agilent 7890A Gas Chromatograph Brochure Manual
- Agilent Two dimensional chromatography: LC-GC online coupling of an Agilent 1260 Infinity LC an Agilent 7890A GC Manual
- Agilent 7890A Network Gas Chromatograph Data Sheet handbook
- Agilent Rapid Forensic Toxicology Screening Using an Agilent 7890A/NPD/5975/DRS GC/MSD System Application Brief
- Agilent Detection of Sulfur Compounds in Natural Gas According to ASTM D5504 with Agilent's Dual Plasma Sulfur Chemiluminescence Detector (G6603A) on the 7890A Gas Chromatograph Manual
- Agilent The Use Of Automated Backflush on the 7890A/5975A GC-MS System Manual
- Agilent High-Pressure Injection Device for the Agilent 7890A 6890 Series Gas Chromatographs Manual
- Agilent Simultaneous Analysis of Trace Oxygenates Hydrocarbons in Ethylene Feedstocks Using Agilent 7890A GC Capillary Flow Technology Manual
- Agilent High-Pressure Liquid Injection Device for the Agilent 7890A 6890 Series Gas Chromatographs Manual
- AGILENT Dual Channel Simulated Distillation of Carbon Sulfur with the Agilent 7890A GC 355 Sulfur Chemiluminescence Detector Application Note
- AGILENT Boiling Point Range of Fatty Acid Methyl Esters Using the 7890A Gas Chromatograph Low Thermal Mass (LTM) System 7693A Tower Tray Application Note
- AGILENT Fast Hydrocarbon Sulfur Simulated Distillation Using the Agilent Low Thermal Mass (LTM) System on the 7890A GC 355 Sulfur Chemiluminescence Detector Application Note
- Agilent Trace-Level Analysis of Melamine in Milk Products on Agilent 7890A/5975C GC/MSD Using a New Agilent J W DB-5ms Ultra Inert Column SampliQ SCX Cartridges Manual
- Agilent Rapid Screening of Melamine Cyanuric Acid in Milk Products Using Agilent J W HP-5ms GC Column Agilent 7890A/5975C GC/MSD with Column Backflushing Manual
- Agilent Comprehensive GC System Based on Flow Modulation for the 7890A GC Manual
- Agilent Analysis of Suspected Flavor Fragrance Allergens in Cosmetics Using the 7890A GC Capillary Column Backflush Manual
- Agilent Maintaining Compound Retention Times with the Backflush Enabled Pressure Controlled Tee Configuration for Agilent 7890A GCs with Agilent 5975 Series MSD Agilent 7000 Series Triple Quadrupo
- Agilent 7890A Chromatographe en phase gazeuse Dépannage Manual
- Agilent 7890A Chromatographe en phase gazeuse Guide de référence rapide
- Agilent 7890A Chromatographe en phase gazeuse Guide d utilisation
- Agilent 7890A Gaschromatograph handbook
- Agilent 7890A GC Sicherheitshandbuch handbook
- Agilent 7890A Gascromatografo Risoluzione dei problemi
- Agilent 7890A Gascromatografo Manutenzione del gascromatografo
- Agilent 7890A Gascromatografo Guida operativa Manual
- Agilent 7890A Gascromatografo Guida alla sicurezza
- Agilent 7890A GC Quick Reference Manual
- Agilent 7890A Cromatógrafo de gases Diagnóstico y resolución de problemas Manual
- Agilent 7890A Cromatógrafo de gases Referencia rápida Manual
- Agilent 7890A Cromatógrafo de gases Guía de funcionamiento
- Agilent G3437A Side-Mounted Thermal Conductivity Detector For the Agilent 7890A Gas Chromatograph Installation Instructions
- Agilent G3510A/G3511A Multimode Inlet for the Agilent 7890A Gas Chromatograph Installation Instructions
- Agilent LTM Series II Rapid Heating/Cooling System for Agilent 7890A GC User Guide
- Agilent 7890A GC Quick Reference
- Agilent 7890A Gas Chromatograph Operating Guide
Questions answered in these manuals
Can I use an uninterruptible power supply (UPS) with the 7890A GC?
No. The manual states that a UPS should not be used with the GC because it can create unsafe conditions.
What type of fuses are required for the 7890A GC?
The instrument uses 20 A, 250 Vac IEC 127 type f fuses for the AC board and 8 A, 250 Vac IEC type fuses for the logic board, and they must be replaced only by trained Agilent service personnel.