Agilent 7890A/5975 Analysis of Flavor/Fragrance Allergens in Cosmetics Using Capillary Column Backflush Application Note

Analysis of Suspected Flavor and Fragrance Allergens in Cosmetics Using the 7890A GC and Capillary Column Backflush Application


Agilent 7890A/5975 Analysis of Flavor/Fragrance Allergens in Cosmetics Using Capillary Column Backflush Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
665 KB
Published
12 June, 2012
Updated
30 September, 2023
Model
7890A
MD5 checksum
889EDC2D2B6EC912D02062E343E1114F

About this manual

The document presents a method for determining flavor and fragrance allergens in cosmetic products using Gas Chromatography-Mass Spectrometry (GC-MS). The primary challenge addressed is the presence of low-volatility matrix constituents, such as detergents and waxes, which can contaminate the analytical system if samples are introduced without selective preparation.

To solve this, the application proposes an alternative method using a standard split/splitless inlet and Capillary Flow Technology with a QuickSwap device at the column outlet. By controlling the column outlet pressure electronically after the target solutes have eluted, remaining sample components are forced back out of the column head into the split vent trap (backflush). This technique significantly reduces column and detector contamination, eliminates ghost peaks, and increases sample throughput by reducing total analysis time from approximately 14 minutes to 8 minutes.

What's inside

  • Introduction
  • GC-MS Conditions
  • Sample Preparation
  • Backflush conditions
  • Results

Frequently asked questions

What are the inlet pressure and auxiliary pressure settings for the backflush?

The inlet pressure is set to 2 psi and the auxiliary pressure is set to 70 psi.

How long must samples be placed in an ultrasonic bath for extraction?

Samples should be placed in an ultrasonic bath for 15 minutes to dissolve the target solutes.