Agilent 7890A Ultra-Fast TPH Analysis with LTM GC and Dual-Tower Injection Application Note

This application note is targeted for ultra-high productivity of total petroleum hydro- carbons (TPH) analysis in environmental laboratories. Agilent's Low Thermal Mass (LTM) technology is employed here to perform ultra-fast gas chromatographic (GC) separations. The LTM technology uses a column module combining a fused silica cap- illary column with heating and temperature-sensing components wound around it, which can be heated and cooled very efficiently. In this application note, the speed of analysis for the hydrocarbon group eluting between C10 and C44 can be dramatically increased to about 13 times faster than a conventional method. In addition, the ultra- fast cooling function of an LTM module can reduce the total GC cycle time to 5.1 minutes. The simultaneous dual-tower injection from Agilent is used to further double productivity. The final result for TPH analysis productivity is 5.1 minutes per two samples.


Agilent 7890A Ultra-Fast TPH Analysis with LTM GC and Dual-Tower Injection Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
487 KB
Published
02 July, 2012
Updated
30 September, 2023
Model
7890A
MD5 checksum
4D55BCCEF312DD6770FD739115384084

About this manual

The document details a method for analyzing total petroleum hydrocarbons (TPH) in environmental laboratories with high speed. It utilizes the Agilent Low Thermal Mass (LTM) system, which employs a specialized column module to heat and cool efficiently, reducing cycle times significantly compared to conventional air bath ovens.

Additionally, the note covers the use of simultaneous dual-tower injection on the Agilent 7890A Gas Chromatograph. This configuration allows for two complete analyses per run using identical channels, effectively doubling lab productivity when combined with the speed gains from LTM technology.

What's inside

  • Introduction
  • Standard Preparation
  • Sample Preparation
  • Instrumentation and Conditions
  • Gas Chromatograph Conditions
  • Ultra-Fast Separation of n-alkanes
  • Simultaneous Dual-Channel Analysis
  • Quantitative Analysis of TPH

Specifications

Compatible ModelAgilent 6890 GC
LTM System Part NumberG6579A
Column Module Length (Standard)5 meters
Minimum Cycle Time (LTM Method)5.1 minutes

Frequently asked questions

Which software is used for data processing?

ChemStation (32-bit version B.04.01) is used for method translation, data generation, and integration.

Can the LTM system be used with other models?

Yes, a version of the LTM system is also available for the Agilent 6890 GC.