AGILENT Fast Hydrocarbon Sulfur Simulated Distillation Using the Agilent Low Thermal Mass (LTM) System on the 7890A GC 355 Sulfur Chemiluminescence Detector Application Note

A fast multielement simulated distillation method (SimDis) based on Low Thermal Mass (LTM) column technology is described.


AGILENT Fast Hydrocarbon Sulfur Simulated Distillation Using the Agilent Low Thermal Mass (LTM) System on the 7890A GC 355 Sulfur Chemiluminescence Detector Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
191 KB
Published
02 July, 2012
Updated
30 September, 2023
Model
7890A
MD5 checksum
3CC822A3BCE05EA2482C2C4F7A148E9A

About this manual

The document details a method for fast hydrocarbon and sulfur simulated distillation (SimDis) utilizing the Agilent Low Thermal Mass (LTM) column technology. This system uses resistive heating to allow rapid temperature programming and extremely fast column cooldown, significantly shortening analytical cycle times compared to conventional air bath GC ovens. The method combines hydrocarbon SimDis with an FID and sulfur SimDis using the Agilent 355 sulfur chemiluminescence detector (SCD) in a series configuration.

What's inside

  • Introduction
  • Experimental
  • Gas Chromatographic Conditions
  • Processing Two-Channel Data
  • Results and Discussion
  • RGO SimDis Results
  • Calibration Mixture Repeatability

Specifications

Column dimensions5 m x 320 mm x 1 pm
Column temperature program45 to 350 °C at 150 °C/min
Run time (LTM)less than 2.5 minutes
Run time (Conventional)15 minutes
Speed improvementat least six times shorter

Frequently asked questions

How much faster is the LTM method compared to conventional ASTM D2887?

The LTM method produces a run time that is at least six times shorter than the conventional ASTM D2887 procedure.

What is the run time for the reference gas oil (RGO) analysis using LTM?

The run time for the RGO analysis using LTM is less than 2.5 minutes.

Does the LTM method agree with ASTM D2887 specifications?

Yes, the SimDis results agree with the specifications in ASTM D2887 for the reference gas oil checkout sample.