Agilent TC-C18(2) and HC-C18(2) Reverse Phase LC Columns Brochure
This document introduces the new Agilent TC-C18(2) and HC-C18(2) reverse-phase HPLC columns, highlighting their superior peak shape, reproducibility, and reliability for routine laboratory analysis. Developed to meet high-quality standards, these columns are designed for a wide range of sample types and mobile phases. This manual is written in Spanish.
Read onlineAgilent HC-C18(2) specifications
| Operating Pressure | 5000 psi (340 bar) |
|---|---|
| Column Dimensions | 4.6 mm id x 12.5 mm |
| Void Volume | < 150 µL |
| Maximum Temperature | 60 °C |
Source: Agilent HC-C18(2) TC-C18(2) High Performance Reliance Cartridge Guard-Columns Manual
All Agilent HC-C18(2) documents by type
Questions answered in these manuals
What is the stated particle size for both the TC-C18(2) and HC-C18(2) columns?
Both the TC-C18(2) and HC-C18(2) columns have a stated particle size of 5 um.
From: Agilent TC-C18(2) and HC-C18(2) Reverse Phase LC Columns Brochure
What is the specified pH range for using the HC-C18(2) column?
The HC-C18(2) column is stable over a stated pH range of 2-9.
From: Agilent TC-C18(2) and HC-C18(2) Reverse Phase LC Columns Brochure
How often should the guard cartridge be replaced?
Replace the guard cartridge when system pressure has increased by 10% from normal or after every 100 to 200 injections.
From: Agilent HC-C18(2) TC-C18(2) High Performance Reliance Cartridge Guard-Columns Manual
What is the recommended procedure for flushing a new guard column?
Flush a minimum of 3 mL of methanol through the assembly, then flush with 10 mL of mobile phase.
From: Agilent HC-C18(2) TC-C18(2) High Performance Reliance Cartridge Guard-Columns Manual