Agilent 7890A LTM System Mineral Oil Analysis Application Note

This article introduces a fast GC-FID method for the analysis of mineral oil in environmental samples using a low thermal mass system. The fast oven temperature programming achieved by the low thermal mass system dramatically reduced the analysis cycle time and increased the sensitivity. The regulated method requirements are met for the environmental analysis of the C10-C40 hydrocarbon fraction in soil and water extracts using splitless injection with an analysis time less than three minutes. Cool-down time is less than two minutes, resulting in an injection-injection cycle time of five minutes. Method performance criteria, including repeatability, linearity and solute discrimination, are presented.


Agilent 7890A LTM System Mineral Oil Analysis Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
315 KB
Published
02 July, 2012
Updated
30 September, 2023
Model
7890A
MD5 checksum
BEACB017C1F456CD0C96779818D3161B

About this manual

This document details a method for analyzing hydrocarbon oil index (HOI) in environmental samples using Gas Chromatography with Flame Ionization Detection (GC-FID). The method utilizes an Agilent 7890A GC system equipped with a Low Thermal Mass (LTM) oven module to significantly reduce analysis time compared to standard procedures.

What's inside

  • Introduction
  • Experimental Conditions
  • GC-FID Conditions
  • Results and Discussion
  • System Suitability
  • Calibration
  • Conclusion

Specifications

Analysis time (LTM)less than three minutes
Cool-down timeless than two minutes
Injection-injection cycle timefive minutes
Column dimensions10 m x 0.32 mm, 0.1 pm
Linearity (r²)> 0.999
Limit of Detection (LOD)< 25 mg/L

Frequently asked questions

What is the total cycle time for the LTM method?

The total injection-injection cycle time is five minutes.

Which Agilent system is used for this analysis?

The analysis is performed on an Agilent 7890A GC System equipped with an LTM column module.

What is the required peak area ratio for C40/C20?

The ratio of the peak areas of the C40-C20 fraction versus the C19-C49 fraction should be between 1.25 and 1.40.