Agilent 7890A GC Application Note for ASTM D5501-09

This application note introduces the use of high-resolution gas chromatography to determine the amount of ethanol and methanol in denatured fuel ethanol. The method is used on an Agilent 7890A GC system, which is coupled to Agilent MultiTechnique ChemStation software. This system offers a complete, automated solution for denatured fuel ethanol analysis.


Agilent 7890A GC Application Note for ASTM D5501-09 - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
224 KB
Published
12 June, 2012
Updated
30 September, 2023
Model
7890A
MD5 checksum
8A47BDBF6005120FBCE4CA9B52BB46E5

About this manual

This document details the setup and operation of an Agilent 7890A Gas Chromatography (GC) system configured to run ASTM Method D5501-09 for the analysis of denatured fuel ethanol. The system utilizes the Agilent 7693 Automatic Liquid Sampler and MultiTechnique ChemStation software to ensure precise measurement of ethanol and methanol content.

The note covers the experimental configuration, including specific hardware options, column specifications, and operating conditions such as temperature programs and flow rates. It demonstrates the system's ability to achieve excellent precision and quantitative transfer of compounds, validating the method's suitability for commercial fuel quality assurance.

What's inside

  • Introduction
  • Experimental
  • Results
  • System Configuration
  • Operating Conditions
  • Calibration
  • Quantification
  • Conclusion

Specifications

GC SystemAgilent 7890A Series GC System
Inlet Option150 psi Split/Splitless Inlet with EPC control
Detector OptionCapillary FID with EPC control
SamplerAgilent 7693 Automatic Liquid Sampler
Column TypePDMS, 150 m x 0.25 mm id x 1.0 um film
Data SystemAgilent MultiTechnique ChemStation rev B.04.01
Inlet Temperature300 °C
Carrier GasHelium at 66 psi
Split Ratio200:1
Sample Size0.5 uL injection

Frequently asked questions

What ASTM method is used for analyzing denatured fuel ethanol?

The document uses ASTM Method D5501-09 for the analysis of denatured fuel ethanol.

What software is used for automated calculations?

The Agilent MultiTechnique ChemStation software performs automated calibration and reporting.

What is the split ratio used in the inlet?

The operating conditions specify a split ratio of 200:1.

Download PDF (224 KB)

More Agilent Manuals


View all Agilent manuals →