Agilent 7890A GC Analysis of Trace 2-Ethylhexyl Nitrate in Diesel Using Chemiluminescence Detector Application Note

This report uses the Agilent 7890A GC system configured with a chemiluminescence detector (NCD) to analyze 2-ethylhexyl nitrate in diesel fuel, and the results show that the NCD can provide the required sensitivity and selectivity.


Agilent 7890A GC Analysis of Trace 2-Ethylhexyl Nitrate in Diesel Using Chemiluminescence Detector Application Note - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
234 KB
Published
12 June, 2012
Updated
30 September, 2023
Model
7890A
MD5 checksum
6252BE55551D77F826342091EA531D32

About this manual

The document details an alternative method to ASTM D4046 for determining alkyl nitrate in diesel fuel. It utilizes an Agilent 7890A GC system configured with a chemiluminescence detector (NCD) to analyze 2-ethylhexyl nitrate (2-EHN). The method offers high sensitivity and selectivity for nitrogen over carbon, allowing for the detection of 2-EHN at low concentrations without interference from hydrocarbon backgrounds.

What's inside

  • Application Brief
  • Experiment
  • Results
  • Highlights
  • References

Specifications

Inlet Temperature250°C
Column TypeHP-5MS
Column Dimensions15 m x 0.32 mm
Column Flow Rate3.9 mL/min
Oven Temperature Program60 °C {2 min), to 280 °C {8 min) at 20 °C/min
NCD Temperature200°C
Detector Pressure77 Torr
Dual Plasma Controller Pressure110 Torr
Burner Temperature905 °C
Hydrogen Flow Rate5 sccm
Oxidant Flow Rate10 sccm

Frequently asked questions

What is the detection range for alkyl nitrate using the ASTM D4046 standard?

The ASTM D4046 standard test method uses spectrophotometry with a detection range of 0.03 to 0.30 volume percent.

What is the concentration of 2-EHN as nitrogen in pump diesel fuel?

The concentration determined for 2-EHN as nitrogen in pump diesel is 1.18 ppm.

What is the concentration of 2-EHN as nitrogen in B20 biodiesel?

The concentration determined for 2-EHN as nitrogen in B20 biodiesel is 18.7 ppm.

Does the NCD method suffer from hydrocarbon interference?

No, the method demonstrates high selectivity for nitrogen over carbon and analyzes nitrogen in diesel without suffering from any hydrocarbon interference.