Agilent 7890A GC Application Note EN 15779

The GC analysis of four common polyunsaturated fatty acid methyl esters (PUFA FAMEs) in algal biodiesel is described using method EN 15779. An Agilent 7890A GC system was configured and calibrated according to the procedure outlined in the method. Two samples of B100 biodiesel made from algae oil were each prepared in duplicate and analyzed according to the conditions set forth in the method. In each sample, the four PUFA FAMEs were chromatographically separated and quantified. The analysis precision was calculated and shown to exceed the specifications of the EN 15779 methods.


Agilent 7890A GC Application Note EN 15779 - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
110 KB
Published
11 June, 2012
Updated
28 September, 2023
Model
7890A
MD5 checksum
0B005101E867AF0AD5C73905A054633E

About this manual

This document details the use of the Agilent 7890A Gas Chromatograph to analyze four common polyunsaturated fatty acid methyl esters (PUFA FAMEs) in biodiesel derived from algae and marine oils. The analysis follows the European Committee for Standardization (CEN) method EN 15779 to measure specific fatty acids that can affect fuel quality.

The note provides specific instrument configurations, including hardware options, column details, and operating conditions such as oven ramps and inlet temperatures. It presents results from duplicate analyses of two algal biodiesel samples, demonstrating the system's precision and repeatability against the specifications defined in the EN 15779 standard.

What's inside

  • Abstract
  • Introduction
  • Experimental
  • Results and Discussion
  • Conclusion
  • References

Specifications

GC SystemAgilent 7890A Series GC
AutoinjectorAgilent 7693 Autoinjector
Column123-7032 DB-Wax Column
Inlet ModeSplit at 50:1 split ratio
Detector Temperature250 °C
Column FlowHydrogen at 1 mL/min

Frequently asked questions

What method is used for the analysis?

The analysis uses method EN 15779, a GC method developed by the European Committee for Standardization.

Which PUFA FAMEs are measured?

The method measures methyl eicosatetraenoate (C20:4), methyl eicosapentaenoate (C20:5), methyl docosapentaenoate (C22:5), and methyl docosahexaenoate (C22:6).

What internal standard is used?

A 1.0 mg/mL solution of methyl tricosanoate (C23:0) in n-heptane is used as an internal standard.