Agilent Maintaining Compound Retention Times with the Backflush Enabled Pressure Controlled Tee Configuration for Agilent 7890A GCs with Agilent 5975 Series MSD Agilent 7000 Series Triple Quadrupo

This note describes how to generate compound retention times that are permanent and universal through Retention Time Locking using the Agilent 7890A GCs, Agilent 5975 series MSD and Agilent 7000 series triple quadrupole MS systems


Agilent Maintaining Compound Retention Times with the Backflush Enabled Pressure Controlled Tee Configuration for Agilent 7890A GCs with Agilent 5975 Series MSD Agilent 7000 Series Triple Quadrupo - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
5.2 MB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
BA1C90E4E4A057760DFBA58634547423

About this manual

The document describes a method using the Pressure Controlled Tee (PCT) configuration to achieve rapid backflushing of gas chromatography columns without venting the mass spectrometer system. This approach prevents sample matrix carryover and extends column life while maintaining analytical robustness.

Furthermore, the note details how Retention Time Locking (RTL) can be applied to this setup. By locking retention times to a specific compound (chlorpyrifos-methyl), analysts can ensure that scheduled events in MS acquisition modes like SIM or MRM remain valid even after column maintenance or replacement. The text provides experimental data showing that retention times are reproduced within 1 second of expected values, simplifying method development and database usage for pesticide analysis.

What's inside

  • Pressure Controlled Tee Configuration
  • Retention Time Locking (RTL)
  • Backflushing without MS Venting
  • Column Maintenance Procedures
  • Ultimate Union Connector Usage
  • Japan Positive List Pesticide Analysis
  • SIM and MRM Mode Stability

Specifications

Locking Compound Retention Time Target13.443 min
Column Configuration (Experimental)Two 15-m DB-5ms columns
Column Flow Difference Setting0.04 mL/min
Retention Time Accuracybetter than an average of 1-sec

Frequently asked questions

How does the Pressure Controlled Tee configuration improve acquisition?

It provides a simple approach to rapid backflushing and GC servicing on all MS systems without venting the mass spectrometer.

What is Retention Time Locking (RTL) used for?

It makes compound retention times permanent and universal, so complex MS acquisition methods do not need to be changed when columns are trimmed or replaced.

How is mechanical retention time locking achieved after column replacement?

By replacing the removed section of the column with an equal length of capillary column using the Ultimate Union connector included in the G1472A kit.