Troubleshooting guide for the Agilent 7890A gas chromatograph, covering symptoms, possible causes, leak checks, and detector tasks.
This manual provides troubleshooting information for the Agilent 7890A gas chromatograph. It is organized around symptoms and the corresponding tasks to perform, with possible causes listed for each problem.
Sections cover ALS and detector symptoms, chromatogram symptoms, GC Not Ready symptoms, shutdown symptoms, GC power and communication symptoms, leak checking, and troubleshooting tasks. It also describes configurable items to keep current and the run, maintenance, and event logs. This manual is written in Chinese.
| NPD output (typical) | 5.0 to 50.0 pA |
|---|---|
| FPD Lit offset (typical) | 2.0 pA |
| Internal log capacity | Up to 250 entries per log |
Look through the vent on the detector cover for an orange-red glow from the rubidium bead.
Remove the rubber dropper from the detector exhaust and place a mirror or shiny object near the aluminum exhaust pipe; steady water condensation indicates the flame is lit.
The typical value is 2.0 pA.
Each internal log can hold up to 250 entries.