A precolumn backflush system based on capillary columns using midpoint pressure control is described. Midpoint backflush is made possible with a Capillary Flow Technology (CFT) purged union controlled by an AUX EPC channel on the Agilent 7890A GC system. The key application discussed is prefractionation of crude oil that provides a high resolution separation of the C4 to C12 cut. A general backflush method using Polywax 500 is presented to illustrate the backflush concept.


Agilent 7890A GC Prefractionator Application Note - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
10
File format
PDF
File size
1.6 MB
Published
02 July, 2012
Updated
29 September, 2023
Model
7890A
MD5 checksum
69C3923FEF4428C9C341910F1E241613

About this manual

The document details a method for analyzing the light ends (C4 to C12) of crude oil and other petroleum fractions using a precolumn backflush system. This technique utilizes a Capillary Flow Technology (CFT) purged union controlled by an AUX EPC channel on the Agilent 7890A GC system to vent heavy components, preventing column contamination and reducing analysis time.

What's inside

  • Introduction
  • Experimental
  • Description
  • Fine tuning backflush time
  • Easily Protect the Analytical Column

Frequently asked questions

What specific fraction of crude oil does this system separate with high resolution?

The system provides a high resolution separation of the C4 to C12 cut of crude oil.

What is the maximum pressure difference allowed between the inlet and auxiliary channels?

Differences less than 0.1 psig must be avoided to assure stable operation.

What is the recommended flow setting range for the precolumn relative to the analytical column?

The precolumn flow should be set between 70% and 85% of the analytical column flow.