Agilent 7890A/7000A Series Triple Quadrupole GC/MS Backflushing Application Note

This document introduces a method for routine detection and quantification of biomarkers in petroleum using the Agilent 7890A/7000A Series Triple Quadrupole GC/MS with backflushing using a Pressure Controlled Tee configuration. In a single run, diverse biomarkers from several transitions can be detected, confirmed, and quantified at levels as low as 2 ppm, with RSDs well below 5%. This method is suitable for “fingerprinting” of petroleum samples and the deconvolution of oil mixtures in complex, multisource petroleum systems.


Agilent 7890A/7000A Series Triple Quadrupole GC/MS Backflushing Application Note - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
10
File format
PDF
File size
1.3 MB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
EA3EBB10CE800418F1CD213C9B205B7B

About this manual

This document presents a method for the routine detection and quantification of biomarkers in petroleum samples using the Agilent 7890A/7000A Series Triple Quadrupole GC/MS. The method utilizes a Pressure Controlled Tee (PCT) configuration with backflushing to remove higher boiling substances from the column prior to subsequent runs. This approach allows for the detection of diverse biomarkers at levels as low as 2 ppm with relative standard deviations (RSDs) well below 5% in a single run. The technique is suitable for fingerprinting petroleum samples and deconvoluting oil mixtures in complex systems.

What's inside

  • Abstract
  • Introduction
  • Standards and Samples
  • Instruments
  • MS SRM Parameters
  • Gas Chromatograph and Mass Spectrometer Conditions
  • Analysis Parameters for Precision Experiments
  • Backflushing with a Pressure Controlled Tee Configuration
  • Results and Discussion

Specifications

Detection limit2 ppm
Precision (RSD)below 5%
Source temperature250 °C
Quadrupole temperature150 °C
Delta EMV70 eV
Acquisition modeSelected reaction monitoring

Frequently asked questions

What is the detection limit for biomarkers using this method?

The method can detect, confirm, and quantify biomarkers at levels as low as 2 ppm.

What is the typical precision of the analysis?

The analysis achieves relative standard deviations (RSDs) well below 5%.

How does backflushing improve the analysis?

Backflushing removes higher boiling substances from the column prior to subsequent runs, reducing accumulated chemical noise and increasing throughput.