Agilent 7890A/NPD/5975C/DRS GC/MSD System Application Brief

This system addresses the challenges faced by laboratories performing toxicology screens on forensic samples by combining fast GC to reduce run time, simultaneous collection of scan, SIM, and NPD data in one shorter run, backflushing to prevent detector fouling, and Deconvolution Reporting Software (DRS) to simplify data interpretation. The system is able to find all the molecules detected by the conventional method in significantly less time.


Agilent 7890A/NPD/5975C/DRS GC/MSD System Application Brief - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
204 KB
Published
02 July, 2012
Updated
29 September, 2023
Model
7890A
MD5 checksum
82C3274220BCD83F4C713E387192CFC1

About this manual

The document details a method for analyzing large numbers of forensic toxicology samples containing complex matrix interferences. The approach combines fast gas chromatography (GC) to reduce run times with simultaneous collection of scan, Selected Ion Monitoring (SIM), and Nitrogen Phosphorus Detector (NPD) data in a single shorter run.

Key features include backflushing to prevent heavy matrix components from fouling detectors and the use of Deconvolution Reporting Software (DRS) to simplify data interpretation. The method identifies up to 278 target compounds, including fentanyl, by utilizing full-spectrum searching against a library, ion ratios in ChemStation, and NPD response.

What's inside

  • Highlights
  • Experimental and Results
  • Gas Chromatograph Conditions
  • Mass Spectrometer Conditions
  • Time Savings Comparison
  • DRS Report

Specifications

Total run time8.5 min
Column length15.0m
Oven voltage240V
NPD temperature310°C

Frequently asked questions

How much time is saved using this method compared to the original?

The cycle time required per sample is reduced by 85%.

What software simplifies data interpretation in dirty samples?

Deconvolution Reporting Software (DRS) simplifies data interpretation, especially in dirty samples.

Download PDF (204 KB)