Agilent Protein 80 Manuals

ManualLib has 5 documents for the Agilent Protein 80: 2 product leaflets, 1 application note, 1 user manual. Every file can be read online or downloaded as a PDF for free.

Product Leaflet 2Application Note 1User Manual 1

Key facts

Brand
Agilent
Model
Protein 80
Documents
5
Languages
EN
Last added
2012-07-05

Start here

Agilent Protein 80 Kit Guide

Agilent Protein 80 Kit Guide

User Manual 32 pages · 1.1 MB · EN

The Agilent Protein 80 Kit is a reagent kit designed for use with the Agilent 2100 Bioanalyzer. It includes Protein Chips, Protein 80 Reagents, and supplies necessary for conducting high-sensitivity protein assays. The kit allows for the analysis of up to 10 samples per chip within a sizing range of 5-80 kDa.

Read online

Agilent Protein 80 specifications

Sizing range Protein 805-80 kDa
Sizing range Protein 23014-230 kDa
Typical sizing resolution10 %
Analysis time30 minutes
Number of samples per chip10 samples/chip
Sample volume required4ul
Kit stability4 months

Source: Agilent Series II Protein 80 and 230 Kits Product Leaflet

All Agilent Protein 80 documents by type

Application Note 1

How to use the product for one specific task, with worked examples.

User Manual 1

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Questions answered in these manuals

What is the basic principle for protein sizing?

Protein size determination is calculated via a standard curve from the ladder, which converts measured migration times of unknown proteins to their sizes.

From: Agilent Protein Sizing Quantitation Technical Overview for Protein 80/230 Kits and 2100 Bioanalyzer

How is relative protein concentration calculated?

Relative concentration is calculated by comparing peak areas against an internal upper marker added in a defined ratio to each sample.

From: Agilent Protein Sizing Quantitation Technical Overview for Protein 80/230 Kits and 2100 Bioanalyzer

What is the main advantage of using these kits over SDS-PAGE?

The kits offer improved data reliability with pre-packaged reagents and standardized assays that yield highly reproducible data.

From: Agilent Series II Protein 80 and 230 Kits Product Leaflet

How long does it take to analyze 10 samples?

Sizing and quantitation of 10 samples takes less than 45 minutes, including sample preparation and data analysis.

From: Agilent Series II Protein 80 and 230 Kits Product Leaflet

What is the advantage of using the Agilent Protein 80 kit over conventional SDS-PAGE?

The method is fast, reproducible, robust, and replaces gel electrophoresis, staining, and densitometric analysis.

From: Agilent 2100 Bioanalyzer and Protein 80 Kit Milk Protein Analysis Application Note

Can the system distinguish between cow, goat, and sheep milk?

Yes, the electrophoretic pattern derived from each species shows a significantly different protein pattern.

From: Agilent 2100 Bioanalyzer and Protein 80 Kit Milk Protein Analysis Application Note