Agilent PL-GPC 50 Plus Poly(styrene-co-butadiene) GPC Application Note

This application note describes a method for investigating the molecular weight distributions of diene rubber copolymers using an Agilent GPC system.


Agilent PL-GPC 50 Plus Poly(styrene-co-butadiene) GPC Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
3
File format
PDF
File size
530 KB
Published
12 June, 2012
Updated
01 October, 2023
MD5 checksum
A25FEC50A756456F514BD74F20626893

About this manual

This document details the analysis of styrene/butadiene rubber copolymers using gel permeation chromatography (GPC) with viscometry. It explains that while light scattering is unsuitable for these copolymers due to non-uniform chemistry, the Universal Calibration technique using a viscometer provides accurate molecular weight distributions independent of column standards.

The analysis was performed on an Agilent PL-GPC 50 Plus integrated system equipped with a differential refractive index detector, an Agilent PL-BV AQ0RT viscometer, and PLgel 5 pm MIXED-C columns. The results demonstrate the system's ability to resolve molecular weight distributions and determine material structure for elastomeric materials.

What's inside

  • Introduction
  • Instrumentation
  • Materials and Reagents
  • Conditions
  • Results and Discussion
  • Conclusion

Specifications

Flow Rate1 mL/min
Temperature40°C
Column Dimensions300 x 7.5 mm
Operating Temperature Rangeambient to 50 °C

Frequently asked questions

Why is viscometry used instead of light scattering for these samples?

Light scattering cannot be used because the samples are copolymers with non-uniform chemistry, which leads to errors in calculations.

What advantage does the Universal Calibration technique offer?

It gives molecular weight that is independent of the standards used in the column calibration.