Agilent PL-GPC 50 Plus Analysis of Polylactide Application Note

This application note describes how to use gel permeation chromatography (GPC) with a viscometer to analyze polylactide (PLA) and poly(lactide-co-glycolide) (PLGA).


Agilent PL-GPC 50 Plus Analysis of Polylactide Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
3
File format
PDF
File size
529 KB
Published
12 June, 2012
Updated
01 October, 2023
MD5 checksum
9FCEA8A9C5F180F447195CFD9B2F39C0

About this manual

The document details the use of Gel Permeation Chromatography (GPC) coupled with viscometry to determine accurate molecular weight distributions for polylactide (PLA) and poly(lactide-co-glycolide) (PLGA). It explains that while light scattering is unsuitable for these copolymers, the Universal Calibration technique using a viscometer provides reliable results independent of column standards.

The specific instrumentation used includes an Agilent PL-GPC 50 Plus integrated system equipped with differential refractive index detection and a PL-BV AQORT viscometer. The analysis was performed on PLgel 5 um MIXED-D columns under specific conditions involving tetrahydrofuran as the eluent at 40°C.

What's inside

  • Introduction
  • Instrumentation
  • Materials and Reagents
  • Conditions
  • Results and Discussion

Frequently asked questions

What is the operating temperature range for the Agilent PL-GPC 50 Plus system?

The system is designed for operation from ambient to 50 °C.

What is the flow rate used for the analysis of polylactide and poly(lactide-co-glycolide)?

The analysis uses a flow rate of 1 mL/min.

What is the temperature condition for the analysis of polylactide and poly(lactide-co-glycolide)?

The analysis is conducted at a temperature of 40°C.