Key facts
- Brand
- Agilent
- Model
- G2570A
- Industry
- Industrial
- Documents
- 2
- Languages
- EN
- Last added
- 2012-06-07
Start here

Agilent G2570A GC/MSD System Site Preparation Manual Addendum
The document serves as an addendum to the 5973N MSD Site Preparation Manual, detailing unique requirements for installing the G2570A GC/MSD system. It describes the principal subsystems, including the MSD-ready 6850A GC and the 5973N diffusion pumped mass spectrometer, emphasizing that communication between the software and GC is not supported.
Key site preparation areas covered include space and weight dimensions for major components, specific electrical requirements such as voltage ranges and power consumption, air conditioning needs regarding temperature and humidity, carrier gas specifications for helium, and laboratory supply considerations. The manual also provides critical safety warnings regarding electrical grounding, hazardous voltages, hot surfaces, and chemical exhaust venting.
Agilent G2570A specifications
| System Components | MSD-Ready 6850A GC, G2629A Control Module, G2577A Diffusion Pump MSD |
|---|---|
| Software Version | G1701CA MSD Productivity software |
| Power Switch Location | Right side in the lower front corner |
Source: Agilent G2570A GC/MSD System Installation and Operation Differences Guide
All Agilent G2570A documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Mounting, wiring and first power-up. Read it before the hardware goes in.
Questions answered in these manuals
What is the document number for this guide?
The publication number for this manual is G2589-90008.
From: Agilent G2570A GC/MSD System Installation and Operation Differences Guide
Is the JetDirect card supported in the MSD-Ready 6850A GC?
No, the use of a JetDirect card is not supported.
From: Agilent G2570A GC/MSD System Installation and Operation Differences Guide
What is the maximum recommended capillary column ID for the G2570A system?
The largest recommended capillary column has a 250 µm ID.
From: Agilent G2570A GC/MSD System Site Preparation Manual Addendum
Does the MSD ChemStation software communicate with the GC?
No, the software will not communicate with the GC and thus will not store GC parameters nor collect data from GC detectors.
From: Agilent G2570A GC/MSD System Site Preparation Manual Addendum