Quick Start Guide 12 pages · 749 KB · EN

This document serves as a quick start guide for integrating the Agilent 1200 Infinity Series Liquid Chromatography system with the Agilent 500 Ion Trap LC/MS mass spectrometer. It outlines the workflow for setting up the systems to automatically inject samples using MS Workstation and ChemStation software.

Read online

All Agilent MS with the Agilent 1200 Infinity Series documents by type

Quick Start Guide 1

The shortest path to a working setup, usually only a few pages.

Questions answered in these manuals

What software is used for automatic sample injection?

The system uses MS Workstation and ChemStation software to perform automated sample injection and analysis.

From: Agilent 500 Ion Trap LC/MS with the Agilent 1200 Infinity Series Quick Start Guide

How do you ensure runtimes are synchronized?

You must make sure the LC program run time in ChemStation is equal to or longer than the MS program run time.