Agilent 7697A Headspace Sampler User Manual
The Agilent 7697A Headspace Sampler is designed to analyze volatile organic compounds using gas chromatography. It captures the headspace volume above a sample matrix where volatile compounds exist in gaseous form. This device is particularly useful when analyzing solid, pasty, or liquid samples that are difficult to inject directly into a GC inlet. This manual is written in German.
Read onlineAgilent 7697A specifications
| AC board fuses (JF1, JF2) | 8 A, 250 VAC, IEC 127 Type F (no time delay), glass body |
|---|---|
| Logic board battery (BT1) | 3 V lithium cell, 0.048 A-HR, Panasonic model BR 1225 |
| Safety standards | CSA C22.2 No. 61010.1, UL 61010-1, IEC 61010-1, EN 61010-1 |
| EMC / RFI | CISPR 11/EN 55011 Group 1 Class A; IEC/EN 61326 |
| Product class | Class 1 safety product with protective grounding |
| Quality system | Designed and produced under an ISO 9001 certified quality system |
All Agilent 7697A documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
- Agilent 7697A Headspace Sampler Safety Manual 14 p
- Agilent 7697A Headspace Sampler User Manual 86 p
- Agilent 7697A Headspace Sampler Getting Started Guide 18 p
- Agilent 7697A Head Space Sampler Safety Guide 16 p
- Agilent 7697A Headspace Sampler User Manual 94 p
- Agilent 7697A Headspace Sampler User Manual 94 p
- Agilent 7697A Head Space Sampler Safety Manual 16 p
- Agilent 7697A Headspace Sampler Getting Started Manual 18 p
Ratings, dimensions and ordering codes, for selection and design work.
How to use the product for one specific task, with worked examples.
Disassembly, diagnostics and part numbers, for repair.
- Agilent Optimizing Vial Pressurization Parameters for the Analysis of USP 467 Residual Solvents Using the 7697A Headspace Sampler
- Agilent Echantillonneur d'espace de tête Agilent 7697A Mise en route
- Agilent Muestreador de espacio de cabeza 7697A de Procedimientos iniciales Manual
- Agilent 7697A Headspace-Probengeber Erste Schritte Manual
- Agilent 7697A Headspace Sampler Troubleshooting
Questions answered in these manuals
Can the Agilent 7697A be connected to a UPS?
No. The manual states not to use an uninterruptible power supply, because continued power during an outage would create an unsafe condition.
Can flammable gas be used to pressurize sample vials?
No. The 7697A does not support pressurizing vials with flammable gases such as hydrogen or argon/methane mixture, as this creates an explosion hazard.
What is the Agilent 7697A Headspace Sampler?
It is a pressure loop headspace sampling system that has a capacity for either 12 vials or 111 vials.
How does the instrument determine if a transmission line is installed?
The program controlling the headspace sampler identifies most hardware, such as detecting whether a transmission line is installed.
What other documents are available for installing and maintaining the Agilent 7697A?
Additional information on installation, operation, maintenance, and troubleshooting can be found on the instrument's included Agilent GC and GC/MS hardware user information and utility DVD.
What specific safety document must be read before using the Agilent 7697A?
Users must read the Safety Manual prior to operating the headspace sampler and whenever necessary.