Agilent 6540 Accurate-Mass Q-TOF Technical Overview

Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF Technical Overview


Agilent 6540 Accurate-Mass Q-TOF Technical Overview - cover page
Brand
Agilent
Document type
Other
Language
English
Pages
8
File format
PDF
File size
178 KB
Published
05 July, 2012
Updated
01 October, 2023
MD5 checksum
B01116D96B9A2BBE688E2707313EFCC0

About this manual

The document presents a technical overview demonstrating how the Agilent 6540 Accurate-Mass Q-TOF LC/MS System can perform metabolic stability, profiling, and identification in a single experiment. This approach maximizes information obtained from one analysis compared to using separate methods for quantitation and identification.

What's inside

  • Abstract
  • Introduction
  • Experimental Instruments
  • Sample Preparation
  • UHPLC Run Conditions
  • OTOF and Triple Quadrupole MS Conditions
  • Results and Discussion
  • Metabolic Stability Results
  • Simultaneous Profiling of Metabolites
  • Reliable Identification of Metabolites

Specifications

Drying Gas Temperature350°C
Sheath Gas Flow12 L/min
Capillary Voltage4000 V
Mass Range (MS and MS/MS)100-1000 m/z
Acquisition Rate (Combined Mode)5 spectra/sec
Collision Energy28 eV

Frequently asked questions

What modes does the Agilent 6540 operate in for this study?

The system was operated in an initial MS-only mode at 5 Hz acquisition rate and an automatic MS/MS mode with expected metabolites on the inclusion list.

How does the Agilent 6540 compare to the Triple Quadrupole system?

The quantitative data obtained using the 6540 QTOF showed good correlation with analysis performed on the Agilent 6460 Triple Quadrupole LC/MS System.

What mass accuracy did the OTOF spectra demonstrate?

The OTOF spectra for buspirone and its metabolites showed mass errors of less than 1 ppm, enabling confident identifications.