This Technical Note describes the performance of the Agilent 7100 CE system using the Cation Solution Kit. The kit contains necessary supplies for the separation of cations.
This document is a technical note describing the use of the Agilent Cation Solution Kit with the Agilent 7100 Capillary Electrophoresis System. The kit contains necessary supplies, including buffers and reagents, ready-to-use for the separation of cations in indirect UV detection mode. It is also suitable for separating low-molecular weight amines.
The note details an experimental method using a DAD detector with a 56-cm effective length capillary. It provides specific settings for the instrument setup, including voltage (30 KV), current (300 pA), and injection parameters. The document also includes data on migration time reproducibility and tap water analysis results.
| Part Number | 5064-8206 |
|---|---|
| Voltage | 30 KV |
| Current | 300 pA |
| Power | 6w |
| Stoptime | 8:00 min |
The part number for the Cation Solutions Kit is 5064-8206.
The Cation Test Sample includes ammonium, potassium, sodium, calcium, and magnesium at 100 ppm each.
The buffer should be stored at 4 °C.