Agilent 6530 Accurate-Mass Q-TOF LC/MS System with Agilent 1290 Infinity LC for Multi Plant Growth Regulator Analysis from Grapes handbook

This document introduces a method for screening 12 plant growth regulators (PGRs) in grape samples using Agilent 6530 accurate-mass Q-TOF LC/MS system with Agilent 1290 Infinity LC. The method shows excellent sensitivity and outstanding mass accuracy.


Agilent 6530 Accurate-Mass Q-TOF LC/MS System with Agilent 1290 Infinity LC for Multi Plant Growth Regulator Analysis from Grapes handbook - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
14
File format
PDF
File size
2.5 MB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
4E807C80133A445033E978CC4C7732D9

About this manual

This document presents a sensitive and effective method for screening 12 Plant Growth Regulators (PGRs) in grape samples using the Agilent 6530 Accurate-Mass Q-TOF LC/MS system coupled with the Agilent 1290 Infinity UHPLC. The method utilizes a simple, inexpensive dispersive C18 sample cleanup technique to remove background matrix, enabling the detection of multiclass PGRs with outstanding mass accuracy (less than 1 ppm for 50 ppb and 100 ppb concentrations).

The analysis involves liquid phase extraction followed by dispersive C18 cleanup, internal standard spiking, UHPLC gradient separation, and detection via Jetstream ESI-LC/MS/MS. The document details specific chromatographic conditions, mass spectrometer settings, and results for both positive and negative ion modes, demonstrating the system's capability to identify compounds using accurate mass and structural elucidation via MS/MS.

What's inside

  • Introduction
  • Sample Preparation
  • LC/MS/MS Method
  • Mass Spectrometer Settings
  • Positive Mode Analysis
  • Negative Mode Analysis
  • Mass Accuracy Results
  • MS/MS Fragmentation

Specifications

Total run time (positive mode)16 min
Total run time (negative mode)6 min
Injection volume2 yL
Column temperature40 °C
Flow rate0.2 mL/min
Mass accuracy (50 ppb/100 ppb)less than 1 ppm
Mass accuracy (5 ppb)less than 5 ppm

Frequently asked questions

What PGRs are detected in positive mode?

The PGRs detected in positive mode are Chlormequat, d4 chlormequat, Daminozide, Zeatin, Kinetin, 6-Benzyl adenine, Indol Butyric acid, Forchlorfenuron, and Paclobutrazol.

What PGRs are detected in negative mode?

The PGRs detected in negative mode are Gibbrelic acid, Abscisic acid, and 2, 4 Dichlorphenoxy acetic acid.

What is the sample preparation process?

The sample is prepared by adding a 5-g amount of homogenized grape sample to 40 mL Methanol: Water (80:20), mixing with C18 ODS SPE Bulk, vortexing, centrifuging, and diluting before injection.