Agilent 6530 Accurate-Mass Q-TOF LC/MS System Application Note

This application note describes a method using high resolution accurate mass (HRAM) LC/MS for the quantitation of cystine and related metabolites in white blood cells. Excellent assay performance was achieved for cystine quantitation in white blood cells (WBCs) using the ultra-high resolving power and mass accuracy of an Agilent 6530 Accurate-Mass Q-TOF LC/MS System. Further, related metabolites were successfully identified and quantitatively profiled. The HRAM LC/MS data acquired in this study can be retrospectively analyzed to search for more metabolites and biomarkers without sample re-injection.


Agilent 6530 Accurate-Mass Q-TOF LC/MS System Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
12
File format
PDF
File size
1.6 MB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
FE020BA38E2BA2B9031CA9766B2D2E5B

About this manual

This document presents a high resolution accurate mass (HRAM) LC/MS approach for quantitating cystine and profiling related metabolites in complex biological samples, specifically white blood cells (WBCs). The method utilizes the ultra-high resolving power and mass accuracy of the Agilent 6530 Accurate-Mass Q-TOF LC/MS System connected to an Agilent 1290 Infinity LC System.

The study compares the performance of this HRAM method against a previously validated multiple reaction monitoring (MRM) method using triple quadrupole mass spectrometry. Results indicate that the HRAM method offers excellent sensitivity, linearity, dynamic range, precision, accuracy, and reproducibility, comparable to the MRM method. Additionally, the accurate mass data allows for the retrospective identification of 44 related metabolites without requiring sample re-injection.

What's inside

  • Abstract
  • Introduction
  • Experimental
  • Instrumentation
  • Results and Discussion
  • Calibration curve linearity and range
  • Precision and accuracy

Specifications

Mass resolving power10K
Ionization modePositive
Drying gas temperature300 °C
Sheath gas temperature400 °C
Capillary voltage3,750 V
Fragmentor voltage200V
Mass range100-1,000 m/z
Acquisition rate2 Hz

Frequently asked questions

What is the limit of quantitation (LOQ) for cystine in WBCs?

The LOQ of cystine in WBCs is 0.02 uM, or 40 fmol on-column.

What is the dynamic range of the cystine calibration curve?

The cystine calibration curve shows excellent linearity in the dynamic range of 0.02 - 4 uM.

Which software is used for data acquisition and analysis?

MassHunter Workstation Software (B.03.01) is used for data acquisition, and MassHunter Quantitative Analysis Software (version B.04.00) is used for quantitation.