Agilent SepTech ST150 10-C18 Peptide Resolution Application Note

This application note presents a method for separating peptides using acetonitrile-free gradient elution on SepTech ST150 10-C18. In response to the global shortage of acetonitrile, alternative solvents such as ethanol have been investigated, and the performance of SepTech ST150 10-C18 is compared to three leading competitor materials. The results demonstrate the superior resolution of peptides achieved by SepTech ST150 10-C18 under acetonitrile-free mobile phase conditions.


Agilent SepTech ST150 10-C18 Peptide Resolution Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
3
File format
PDF
File size
721 KB
Published
11 June, 2012
Updated
29 September, 2023
MD5 checksum
65DC21623E7D2910A01C5EF34CCCED01

About this manual

Agilent Technologies investigated alternative solvents for reversed phase HPLC analysis of peptides due to global shortages in acetonitrile. This document evaluates ethanol as a viable substitute, comparing its performance against acetonitrile and three leading competitor columns.

The study utilizes the SepTech ST150 10-C18 column with a mixture of four peptides (Oxytocin, Angiotensin II, Angiotensin I, and Insulin). Results indicate that while ethanol provides similar solvent strength to acetonitrile, it alters selectivity, causing partial co-elution between insulin and angiotensin I. Additionally, ethanol exhibits higher UV absorbance at 220 nm compared to acetonitrile.

What's inside

  • Introduction
  • Materials and Reagents
  • Sample Preparation
  • Reference Chromatogram
  • Results
  • Competitor Column Comparison Conclusion

Frequently asked questions

What is the stated particle size of the SepTech ST150 10-C18 column?

The column has a particle size of 10 micrometers.

What is the recommended flow rate for the peptide separation described?

The procedure specifies a flow rate of 1.0 mL/min.

Download PDF (721 KB)