Agilent ZORBAX Eclipse Plus C18 Improved Simvastatin Analysis Application Note

This document describes how to use Agilent ZORBAX Eclipse Plus C18 5µm, 3.5µm and 1.8µm columns for simvastatin analysis


Agilent ZORBAX Eclipse Plus C18 Improved Simvastatin Analysis Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
12
File format
PDF
File size
536 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
18DBEAC956566A13C0AF854FB815B6E2

About this manual

Simvastatin is a widely prescribed cholesterol-lowering drug. This document details how to modify the standard USP method for Simvastatin tablet analysis to utilize shorter Agilent ZORBAX Eclipse Plus C18 columns. The modifications allow for significantly reduced analysis times and solvent usage while maintaining required performance.

What's inside

  • Introduction
  • Experimental
  • Sample and Mobile Phase Preparation
  • Chromatographic Performance
  • Robustness Testing
  • Column Batch Reproducibility

Specifications

Column Length (USP Method)250 mm
Run Time (USP Method)10 min
Flow Rate1.5 mL/min
Column Temperature45 °C
Detection Wavelength238 nm
Injection Volume (250 mm column)10 uL
Injection Volume (100 mm column)6 uL
Injection Volume (50 mm column)2 uL

Frequently asked questions

What is the mobile phase composition?

The mobile phase is a mixture of 65% acetonitrile and 35% phosphate buffer at pH 4.5.

What is the standard run time for the USP method described in this note?

The original USP method has a 10-minute run time.

What is the maximum pressure limit for running these columns?

All columns are run at pressures under 400 bar.