Agilent Eclipse Plus C18 Chromatography of Nitrogen-Containing Compounds Application Note

This work demonstrates that Eclipse Plus C18 can be used for high-performance liquid chromatography (HPLC) separation of difficult to separate nitrogen-containing compounds without the use of triethylamine (TEA), which provides a new method for separating such compounds. In addition, the low end-group activity of Eclipse Plus C18 makes it a preferred HPLC column for analysis of pharmaceuticals and other amine-containing compounds.


Agilent Eclipse Plus C18 Chromatography of Nitrogen-Containing Compounds Application Note - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
167 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
8E68C936275AE3C54E4E94B138B92549

About this manual

This document addresses peak tailing issues in reversed-phase HPLC when analyzing basic and nitrogen-containing compounds, such as pharmaceuticals. It explains that traditional solutions like lowering pH, increasing buffer strength, or adding triethylamine (TEA) can introduce selectivity changes, resolution issues, or column contamination.

The text introduces the Agilent Eclipse Plus C18 column as a superior alternative. Due to its proprietary silica treatment and improved bonding, this column minimizes silanol activity, allowing for excellent peak shape and efficiency for amine-containing analytes without the need for mobile phase modifiers like TEA.

What's inside

  • Introduction
  • Application
  • Experimental
  • Results and Discussion
  • Conclusions
  • References

Specifications

Column Dimensions (Long)4.6 x 75 mm
Particle Size3.5 um
Part Number (Long)959933-902
Part Number (Short)959963-902

Frequently asked questions

Why is triethylamine used in HPLC analysis?

Triethylamine acts as a competing base to reduce the availability of stationary phase silanols, which reduces peak tailing for basic compounds.

What are the disadvantages of using triethylamine?

It adds an additional component to the mobile phase, alters the column in a way not easily reversed, and can cause peak tailing issues in LC-MS.

How does the Eclipse Plus C18 column improve peak shape?

It utilizes a proprietary silica treatment and improved bonding to minimize silanol activity, providing excellent peak shape without TEA.