Agilent ZORBAX Eclipse Plus C18 and Phenyl Columns Application Note

This application note describes the use of Agilent ZORBAX Eclipse Plus C18, Eclipse Plus Phenyl-Hexyl, Eclipse XDB-Phenyl and StableBond SB-Phenyl Rapid Resolution columns to analyze nine common food additives using an isocratic solvent system. The mobile phases used include methanol or acetonitrile with 20 mM phosphate buffer at pH 2.5 or 0.2 percent formic acid. While excellent peak shape is found on all columns examined, selectivity and resolution differences are found when comparing C18 and various phenyl columns, especially the Eclipse Plus Phenyl-Hexyl and StableBond SB-Phenyl columns. In addi- tion, selectivity differences and peak shape are preserved even when using a formic acid mobile phase modifier. The use of formic acid would allow verification of unknown materials using mass spectrometry.


Agilent ZORBAX Eclipse Plus C18 and Phenyl Columns Application Note - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
445 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
D5BEFF9086AAE4A13575FDE5EC48B1AA

About this manual

This document evaluates the performance of four Agilent ZORBAX columns—Eclipse Plus C18, Eclipse Plus Phenyl-Hexyl, Eclipse XDB-Phenyl, and StableBond SB-Phenyl—for the analysis of nine common non-nutritive food additives. The study utilizes a Rapid Resolution LC system with isocratic solvent systems containing methanol or acetonitrile and phosphate buffers or formic acid.

What's inside

  • Introduction
  • Experimental
  • Results and Discussion
  • Selectivity Differences
  • Mobile Phase Comparison
  • Consumer Product Analysis

Specifications

Column Dimensions4.6mm x 150 mm
Particle Size3.5 microns
Flow Rate1.5 mL/min
Column Temperature25 °C
Injection Volume2.0 µL
Detection Wavelength220 nm

Frequently asked questions

Which column provides the best separation for non-nutritive food additives?

The Eclipse Plus Phenyl-Hexyl column provides the best separation for this application under the given conditions, with a minimum resolution of R = 2.32.

Can formic acid be used instead of phosphate buffer for mass spectrometry?

Yes, the formic acid mobile phase can act as a supplementary mobile phase and would be ideal if MS detection were required.

What is the effect of using acetonitrile instead of methanol?

Using acetonitrile results in significant compression of the phenyl chromatograms and a loss of selectivity differences among the C18 and phenyl columns.