Sensitive ESI-LC/MS/MS Analysis of Dansyl Derivatized Phytoestrogens on Agilent ZORBAX RRHT Eclipse Plus C18 1.8 µm Colu

This article introduces an ESI-LC/MS/MS analysis method of dansyl derivatized phytoestrogens on an Agilent ZORBAX RRHT Eclipse Plus C18 1.8µm column. This method has high sensitivity and accuracy, and can be used to analyze a variety of phytoestrogens.


Sensitive ESI-LC/MS/MS Analysis of Dansyl Derivatized Phytoestrogens on Agilent ZORBAX RRHT Eclipse Plus C18 1.8 µm Colu - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
713 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
4A42D12BA4897E8EFD4D4E89E0DD0B60

About this manual

Phytoestrogens are naturally occurring plant compounds with potential health benefits. This document details a method for analyzing these compounds by derivatizing them with dansyl chloride to enhance sensitivity in LC/MS/MS analysis compared to previous APCI methods.

The study utilizes an Agilent 1200 Series Rapid Resolution LC system coupled with an Agilent 6410 Triple Quadrupole Mass Spectrometer. The method employs a specific ammonium acetate/acetonitrile gradient on an Agilent ZORBAX RRHT Eclipse Plus C18 1.8 µm column to achieve rapid separation and high sensitivity, achieving signal-to-noise ratios greater than 10 for the lowest calibrator.

What's inside

  • Authors
  • Abstract
  • Introduction
  • Experimental
  • Gradient Program
  • MS/MS method development
  • Results and Discussion
  • Calibration curves

Frequently asked questions

What is the flow rate used in the LC/MS analysis?

The flow rate was set to 0.5 mL/min.

What temperature was the column compartment set to?

The thermostatted column compartment was set to 40 °C.

How long were the vials heated during the derivatization step?

The vials were heated at 60 °C for 10 minutes.