Agilent ZORBAX Eclipse Plus C18 Application Note for Simvastatin Analysis

This document describes the improvement of simvastatin analysis performance using Agilent ZORBAX Eclipse Plus C18 columns with different particle sizes. By making simple adjustments, the USP method can be adapted to Rapid Resolution Eclipse Plus C18 3.5µm columns or Rapid Resolution HT Eclipse Plus C18 1.8µm columns, resulting in shorter analysis time and significant solvent savings.


Agilent ZORBAX Eclipse Plus C18 Application Note for Simvastatin Analysis - cover page
Brand
Agilent
Document type
Application Note
Language
Japanese
Pages
12
File format
PDF
File size
753 KB
Published
06 July, 2012
Updated
28 September, 2023
MD5 checksum
10C4A079894707FEA86159F9DC307419

About this manual

This document details the optimization of a USP method for analyzing simvastatin tablets. The original method utilizes an Agilent ZORBAX Eclipse Plus C18 5 µm column (250 mm x 4.6 mm) with a 10-minute analysis time.

The note demonstrates how to modify this method using smaller particle sizes and shorter columns, specifically the Rapid Resolution Eclipse Plus C18 (3.5 µm) and Rapid Resolution HT Eclipse Plus C18 (1.8 µm). These modifications significantly reduce analysis time and solvent usage while maintaining performance requirements. This manual is written in Japanese.

What's inside

  • Introduction
  • Simvastatin Structure
  • HPLC System Configuration
  • Column Specifications
  • Sample Preparation
  • Mobile Phase Preparation
  • USP Method Requirements
  • Chromatographic Performance

Specifications

Original Column Part Number959990-902
Rapid Resolution Column Dimensions4.6 mm x 100 mm, 3.5 µm
Flow Rate1.5 mL/min
Column Temperature45 °C
Detection Wavelength238 nm

Frequently asked questions

What is the analysis time for the original USP method?

The original USP method requires a 10-minute analysis time.

What is the required flow rate for the Simvastatin analysis method?

The method requires a flow rate of 1.5 mL/min.

At what temperature should the column be maintained during analysis?

The column must be kept at a constant temperature of 45 C.

What is the maximum pressure limit for using these columns?

All listed columns can be used with pressures under 400 bar.

Download PDF (753 KB)