Agilent Detector linearity testing for Aripiprazole quality control with the Agilent 1120 Compact LC Agilent TC C18(2) columns Manual

This document describes the detector linearity testing for Aripiprazole quality control using the Agilent 1120 Compact LC and Agilent TC C18(2) columns. The experimental results demonstrate excellent retention time and peak area precision, as well as outstanding linearity of the system.


Agilent Detector linearity testing for Aripiprazole quality control with the Agilent 1120 Compact LC Agilent TC C18(2) columns Manual - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
176 KB
Published
02 July, 2012
Updated
28 September, 2023
MD5 checksum
3B19364F35B667278CAF7EE79938550C

About this manual

This document presents an application note detailing the evaluation of the Agilent 1120 Compact LC system's detector linearity and precision. The study utilizes the antipsychotic drug Aripiprazole as a test compound to assess performance across a wide concentration range of approximately 150 ng to 5,000 ng.

The analysis was conducted using an Agilent TC C18(2) column with a variable wavelength detector (VWD) set to 254 nm. The results demonstrate excellent linearity with a coefficient of correlation greater than 0.9999, retention time precision (RSD) less than 0.07%, and area precision (RSD) less than 0.25%.

What's inside

  • Abstract
  • Introduction
  • Equipment
  • Chromatographic parameters
  • Sample preparation
  • Results and discussion
  • Conclusion

Specifications

Column dimensions150 x 4.6 mm
Particle size5 pm
Flow rate1.0 mL/min
Injection volume5 pL
Column oven temperature40 °C
Detector wavelength254 nm
Area precision (RSD)< 0.25 %
Linearity correlation> 0.9999

Frequently asked questions

What is the concentration range tested for detector linearity?

The study covers a wide range of approximately 150 ng to 5,000 ng.

What is the coefficient of correlation for the linearity test?

The observed linearity correlation was r-squared greater than 0.9999.

Which column was used for the Aripiprazole analysis?

An Agilent TC C18(2) column with dimensions 150 x 4.6 mm and 5 pm particle size was used.