Agilent System suitability testing for Aripiprazole quality control with the Agilent 1120 Compact LC Agilent TC C18(2) columns Manual

The document introduces the system suitability testing for Aripiprazole quality control using Agilent 1120 Compact LC and Agilent TC C18(2) columns. The system provides highly precise retention times and peak areas, as well as low detection limits, making it suitable for drug analysis.


Agilent System suitability testing for Aripiprazole quality control with the Agilent 1120 Compact LC Agilent TC C18(2) columns Manual - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
180 KB
Published
02 July, 2012
Updated
29 September, 2023
MD5 checksum
718497239E4CA45664287FA8F9CA43FD

About this manual

This document serves as an application note detailing the use of the Agilent 1120 Compact LC system for system suitability testing (SST) of Aripiprazole and its precursors in the pharmaceutical industry. It highlights the system's reliability and precision for analyzing drugs and impurities.

The note outlines the experimental setup, including the specific Agilent TC C18(2) column and EZChrom Elite Compact Compliance software used. It presents chromatographic parameters, acceptance criteria based on ICH recommendations, and results from a system suitability test involving replicate injections to verify retention time, peak area, resolution, and other critical parameters.

What's inside

  • System Suitability Testing
  • Aripiprazole Analysis
  • Agilent 1120 Compact LC
  • TC C18(2) Column
  • Acceptance Criteria
  • Chromatographic Parameters
  • Pharmaceutical QA/QC
  • Impurity Analysis

Specifications

Flow rate1.0 mL/min
Column oven temperature40 °C
Detection wavelength254 nm

Frequently asked questions

What are the acceptance criteria for retention time precision?

The relative standard deviation (RSD) of retention time must be less than 0.07%.

What is the limit for peak area precision?

The RSD of area must be less than 1.00%.

What is the minimum required resolution?

The resolution must be greater than 2.00.

Download PDF (180 KB)