Agilent Injector linearity testing for Tramadol quality control with the Agilent 1120 Compact LC Agilent TC C18(2) columns Application Note

This Application Note demonstrates the excellent linearity performance of the Agilent 1120 Compact LC autosampler. The system can be used to perform linearity calibration for samples ranging from 1 to 100 µL. The results show that the Agilent 1120 Compact LC autosampler has a linearity coefficient of R2 > 0.9999.


Agilent Injector linearity testing for Tramadol quality control with the Agilent 1120 Compact LC Agilent TC C18(2) columns Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
541 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
24F5B2AA1EF1843B594C59F014BD53C0

About this manual

The document evaluates the autosampler precision and linearity of the Agilent 1120 Compact LC system. It details a study where Tramadol was injected at varying volumes ranging from 1 µL to 100 µL to test performance across a wide analytical range.

Results indicate excellent retention time precision with an RSD < 0.07% and peak area precision with an RSD < 0.5%. The injector linearity showed a coefficient of correlation (R²) greater than 0.9999, confirming the system's suitability for pharmaceutical QA/QC.

What's inside

  • Abstract
  • Introduction
  • Equipment
  • Sample preparation
  • Chromatographic parameters
  • Sequence table
  • Results and discussion
  • Conclusion

Frequently asked questions

What is the range of injection volumes tested for linearity?

The study covers a wide range from 1 µL to 100 µL injected.

What was the concentration of Tramadol used in the test?

The solution concentration used was 100 ng/µL.

What is the coefficient of correlation for the injector linearity?

The observed linearity correlation (R²) is greater than 0.9999.

Download PDF (541 KB)