Agilent 6890Plus/5973N Nitrofen Determination Application Note

This document introduces Agilent's GC/MSD-NCI technology, which can be used to detect Nitrofen in meat.


Agilent 6890Plus/5973N Nitrofen Determination Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
German
Pages
4
File format
PDF
File size
151 KB
Published
06 July, 2012
Updated
01 October, 2023
MD5 checksum
B287F0F352951A75D42A2FA6E7EA72AE

About this manual

This document describes an analytical method for determining the herbicide Nitrofen in turkey meat using Gas Chromatography/Mass Spectrometry (GC/MS). The application focuses on the advantages of Negative Chemical Ionization (NCI) mode over Electron Impact (EI) mode for detecting analytes with high electron affinity in complex food matrices.

The text details the experimental setup, including the use of Agilent 6890Plus and 5973N MSD instruments with EPC modules and Retention Time Locking (RTL). It outlines the sample preparation procedure according to DFGS 19 and presents results comparing EI and NCI detection modes, highlighting the superior sensitivity and selectivity of NCI for Nitrofen detection. This manual is written in German.

What's inside

  • Keywords
  • Introduction
  • Negative Chemical Ionization
  • Pulsed Splitless Injection
  • Retention Time Locking
  • Experimental Execution
  • Sample Preparation
  • GC/MSD Analysis
  • Instrumentation
  • Results

Specifications

Gas ChromatographAgilent 6890Plus
Mass SpectrometerAgilent 5973N, MSD, CL-Option
Injection Volume1uL, Pulsed Pressure Splitless
NCI Moderator GasMethan
Ion Source Temperature150°C

Frequently asked questions

Which detection mode is preferred for Nitrofen?

Negative Chemical Ionization (NCI) mode is preferred because Nitrofen shows high electron affinity.

What is the detection limit for Nitrofen in a turkey meat matrix under the specified NCI SIM conditions?

The detection limit for Nitrofen in the turkey meat matrix is 200 fg.

What is the retention time deviation of the Nitrofen analyte in the spectrum library compared to the matrix sample?

The retention time deviation is +6 seconds.

What is the concentration range used to create the calibration function for Nitrofen SIM measurements?

The calibration function was created in a concentration range from 50 fg/µL to 100 pg/µL.