EXTECH DO600 Manuals

ManualLib has 3 documents for the EXTECH DO600: 2 user manuals. Every file can be read online or downloaded as a PDF for free.

User Manual 2

Key facts

Brand
EXTECH
Model
DO600
Industry
Industrial
Documents
3
Languages
EN
Last added
2023-10-18

Start here

Extech DO600 Dissolved Oxygen Meter User Guide

Extech DO600 Dissolved Oxygen Meter User Guide

User Manual 12 pages · 282 KB · EN

This user guide explains how to operate the ExStik® DO600 dissolved oxygen meter, including electrode preparation, powering on, calibration, and unit selection. It details features like auto power-off, temperature compensation, and memory storage for up to 25 readings. The guide also covers troubleshooting and battery replacement procedures.

The meter displays dissolved oxygen in % saturation, mg/l, or ppm, and temperature in °C or °F. Calibration is recommended daily, and special steps are provided for zero oxygen calibration and electrode maintenance.

Read online

EXTECH DO600 specifications

Memory Capacity25 points
Power Sourcefour (4) CR2032 Lithium lon Batteries
Auto Power Off Time10 minutes
Polarization Periodthree minute (approx.)
Salinity Compensation Range0 to 50 ppt
Altitude Compensation Range20,000ft

Source: EXTECH DO600 Dissolved Oxygen Meter User's Guide

All EXTECH DO600 documents by type

User Manual 2

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Questions answered in these manuals

How do I calibrate the DO600 meter?

Turn the meter ON, press and hold the MODE/HOLD button until % is displayed, then press and hold the CAL/RECALL button until CAL is shown. The meter will blink '101.7' and show 'SA' before returning to normal mode.

From: Extech DO600 Dissolved Oxygen Meter User Guide

What happens if the meter is not used for more than seven days?

A 3-minute waiting period is required before a measurement can be made, as the electrode needs to re-polarize.

From: Extech DO600 Dissolved Oxygen Meter User Guide

How long does the startup polarization period take?

A three minute (approx.) polarization period must elapse before measurements can be made.

From: EXTECH DO600 Dissolved Oxygen Meter User's Guide

How often should calibration be performed?

Calibration should be performed on a daily basis.

From: EXTECH DO600 Dissolved Oxygen Meter User's Guide

What is the measurement range for dissolved oxygen concentration?

The meter displays concentration from 0 to 20.00ppm (mg/L).

From: Extech DO600 Dissolved Oxygen Meter Product Specifications

Does the DO600 support altitude compensation?

Yes, it features adjustable altitude compensation ranging from 0-20,000ft.

From: Extech DO600 Dissolved Oxygen Meter Product Specifications