ABB NGC8200/PGC1000 Gas Chromatograph User Manual
ABB publishes the NGC8200 and PGC1000 gas chromatograph user manual under document number 2101510MNAH. The manual begins with product and hardware descriptions, including the cast aluminum enclosure, feed-through assembly, analytical module, digital controller assembly with VGA display, and termination board.
It then covers installation, wiring, startup with PCCU32, calibration, network and security configuration, maintenance, and troubleshooting. Troubleshooting includes startup diagnostics, alarm definitions, alarm tests, power, and serial communications.
ABB NGC8200 specifications
| Carrier Gas Purity | 99.995% or higher |
|---|---|
| Sample Line Length (with conditioner) | Up to 50 feet |
| Sample Line Length (without conditioner) | Up to 10 feet |
| Carrier Gas Pressure Setting | 90 PSIG |
| Sample Probe Pressure Setting | 15 PSIG |
| 12V System Min Voltage (J1) | 11.5 VDC |
| 24V System Min Voltage (J1) | 25 VDC |
Source: ABB NGC8200 Measurement and Control System Start-Up Guide
All ABB NGC8200 documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
What software is used for the user interface?
The manual identifies the user interface as PCCU32.
Where can PCCU32 be downloaded?
From the ABB website.
What is the recommended tubing material for the instrument?
Only high-quality all-stainless steel chromatographic grade transmission tubes should be used for carrier, calibration, and sample lines.
From: ABB NGC8200 Measurement and Control System Start-Up Guide
What happens if the instrument detects a leak?
If a leak occurs in the carrier, sample, or calibration gas lines, the equipment will produce non-compliant results.
From: ABB NGC8200 Measurement and Control System Start-Up Guide