This document is a start-up guide for the NGC8200 Measurement and Control System from ABB Inc., Totalflow Products. The guide provides basic information on the installation and start-up of the product, as well as installation instructions for optional equipment. It is recommended that inexperienced technicians follow the installation procedures as described in the guide.


ABB NGC8200 Measurement and Control System Start-Up Guide - cover page
Brand
ABB
Category
Industrial Robot
Document type
User Manual
Language
Chinese
Pages
12
File format
PDF
File size
1.5 MB
Published
02 November, 2022
Updated
28 September, 2023
MD5 checksum
9A11424C0384DB8BBCE1736CA8B6F5C5

About this manual

This document is a Chinese quick-start guide for the ABB Totalflow NGC8200 Measurement and Control System. It is intended for typical installations and recommends experienced technicians consult the full User Manual for detailed information. The guide advises reading it before starting installation to understand the provided procedures.

The guide details the physical installation steps, including finding a suitable location near the sample probe, mounting the device on various surfaces (wall, shelf, pipe), and installing the sample conditioning module and probe. It specifies requirements for tubing materials (stainless steel), gas purity, and pressure settings for carrier, calibration, and sample lines. It also covers connecting ventilation, power supply, and installing the PCCU32 software. This manual is written in Chinese.

What's inside

  • Introduction
  • Basic Installation
  • Sample Conditioning
  • Probe Installation
  • Gas Connections
  • Power Installation
  • Software Installation

Specifications

Carrier Gas Purity99.995% or higher
Sample Line Length (with conditioner)Up to 50 feet
Sample Line Length (without conditioner)Up to 10 feet
Carrier Gas Pressure Setting90 PSIG
Sample Probe Pressure Setting15 PSIG
12V System Min Voltage (J1)11.5 VDC
24V System Min Voltage (J1)25 VDC

Frequently asked questions

What is the recommended tubing material for the instrument?

Only high-quality all-stainless steel chromatographic grade transmission tubes should be used for carrier, calibration, and sample lines.

What happens if the instrument detects a leak?

If a leak occurs in the carrier, sample, or calibration gas lines, the equipment will produce non-compliant results.

What software needs to be installed on the device?

The PCCU32 software located on the included CD must be installed on the device.