ABB NGC8200/PGC1000 Gas Chromatograph User Manual
ABB publishes the NGC8200 and PGC1000 gas chromatograph user manual under document number 2101510MNAH. The manual begins with product and hardware descriptions, including the cast aluminum enclosure, feed-through assembly, analytical module, digital controller assembly with VGA display, and termination board.
It then covers installation, wiring, startup with PCCU32, calibration, network and security configuration, maintenance, and troubleshooting. Troubleshooting includes startup diagnostics, alarm definitions, alarm tests, power, and serial communications.
ABB PGC1000 specifications
| Cycle time | 330 seconds |
|---|---|
| Carrier gas | Helium |
| Carrier pressure | 20 psig |
| Sample size | 40 ul |
| Target retention time | 290 seconds |
| Oven temperature | 60° C |
| Stream Input pressure | 3-15 psig |
All ABB PGC1000 documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
Can hydrogen carrier gas be used with a PGC1000 built for helium?
No, hydrogen carrier gas cannot be used in a PGC1000 built for helium.
What is the maximum tolerance for water in the BBC train?
The maximum tolerance for water in the BBC train is 0.005 mol%.
What software is used for the user interface?
The manual identifies the user interface as PCCU32.
Where can PCCU32 be downloaded?
From the ABB website.