Agilent 7697A Headspace Sampler Safety Manual

Safety manual for the Agilent 7697A Headspace Sampler covering safety warnings, hydrogen safety, fuses, batteries, certifications and recycling.


Agilent 7697A Headspace Sampler Safety Manual - cover page
Brand
Agilent
Category
Analyzers
Document type
User Manual
Language
Chinese
Pages
14
File format
PDF
File size
615 KB
Published
10 November, 2026
Updated
10 November, 2026
Model
7697A
MD5 checksum
13E62E8F0C771BFB4B557FC9E6294E42

About this manual

This is the safety manual for the Agilent 7697A Headspace Sampler, published by Agilent Technologies as part number G4556-97012 (third edition, May 2018). It describes the safety precautions to follow when working with the instrument, including warnings about hazardous voltage, hot surfaces, static discharge and the handling of chemicals.

The manual explains hydrogen carrier gas safety and the instrument's hydrogen safety shutdown behavior, lists the fuses and battery used and who may replace them, and summarizes the safety and regulatory certifications of the 7697A. It ends with sections on proper use, cleaning and recycling of the product. This manual is written in Chinese.

What's inside

  • Headspace sampler safety
  • Important safety warnings
  • Hydrogen safety
  • Fuses and battery
  • Safety and regulatory certifications
  • Proper use
  • Cleaning
  • Recycling products

Specifications

AC board fuses (JF1, JF2)8 A, 250 VAC, IEC 127 Type F (no time delay), glass body
Logic board battery (BT1)3 V lithium cell, 0.048 A-HR, Panasonic model BR 1225
Safety standardsCSA C22.2 No. 61010.1, UL 61010-1, IEC 61010-1, EN 61010-1
EMC / RFICISPR 11/EN 55011 Group 1 Class A; IEC/EN 61326
Product classClass 1 safety product with protective grounding
Quality systemDesigned and produced under an ISO 9001 certified quality system

Frequently asked questions

Can the Agilent 7697A be connected to a UPS?

No. The manual states not to use an uninterruptible power supply, because continued power during an outage would create an unsafe condition.

Can flammable gas be used to pressurize sample vials?

No. The 7697A does not support pressurizing vials with flammable gases such as hydrogen or argon/methane mixture, as this creates an explosion hazard.

What fuse should be used when replacing the line fuse?

Only fuses of the same type and rating may be used, and the product must be disconnected from the mains power before replacing the fuse.

Who may replace the fuses and battery?

Only trained Agilent service personnel may handle the fuses and battery.