Key facts
- Brand
- ABB
- Model
- FEP300
- Category
- Other products
- Industry
- Industrial
- Documents
- 2
- Languages
- IT · EN
- Last added
- 2026-11-14
Start here

ABB ProcessMaster FEP300 Electromagnetic Flowmeter Technical Specification
ProcessMaster FEP300 is an electromagnetic flowmeter for process industry applications. The document describes the compact ProcessMaster 300 version with basic functionality, including intuitive operation, non-contact programming keys, Easy Set-up, NAMUR status messages, help texts on the display, and diagnostic functions such as ScanMaster.
It lists compact models, nominal diameters from DN 3 to 2000, process flange connections, lining and electrode materials, IP65 or IP67 protection, fluid temperatures from -25 to 180 °C, power supply options, current, pulse and switch outputs, HART as standard communication with optional PROFIBUS PA or FOUNDATION Fieldbus, and Ex and pressure equipment approvals. This manual is written in Italian.
ABB FEP300 specifications
| Measurement accuracy | Standard 0.4% of measured value; optional 0.2% of measured value |
|---|---|
| Nominal diameter | DN 3 to 2000 (1/10 in to 80 in) |
| Process connection | Flange according to DIN 2501 / EN 1092-1, ASME B16.5 / B16.47, JIS, AS2129 |
| Nominal pressure | PN 10 to 100; ASME CL 150, 300, 600, 900, 1500, 2500 |
| Lining materials | Hard rubber (DN 15 to 2000), soft rubber (DN 50 to 2000), PTFE (DN 10 to 600), PFA (DN 3 to 200), ETFE (DN 25 to 600), Linatex (DN 50 to 600) |
| Conductivity | Greater than 5 µS/cm (20 µS/cm for demineralized water) |
| Electrode materials | Stainless steel, Hastelloy B, Hastelloy C, platinum-iridium, tantalum, titanium, tungsten carbide |
| Process connection material | Steel or stainless steel |
| Protection rating | IP65 or IP67 |
| Fluid temperature | -25 to 180 °C (-13 to 356 °F) |
| Power supply | 100 to 230 V AC (-15/+10%), 24 V AC (-30/+10%), 24 V DC (-30/+30%) |
| Current output | 4 to 20 mA, active or passive |
| Pulse output | Active or passive, adjustable on site via software |
| Switch output/input | Optocoupler with programmable function |
| Display | Graphic display, adjustable |
| Housing | Compact form, single-chamber or dual-chamber |
| Communication | HART standard; PROFIBUS PA and FOUNDATION Fieldbus optional |
| Ex approvals | ATEX/IECEx Zone 1, 2, 21, 22; FM/cFM Class 1 Div. 1 (up to DN 300) or Div. 2; NEPSI Zone 1, 2; GOST Zone 1, 2 |
| Pressure equipment directive | 97/23/EC, conformity assessment category III, fluid group 1 |
| CRN | On request |
Source: ABB ProcessMaster FEP300 Electromagnetic Flowmeter Technical Specification
All ABB FEP300 documents by type
Ratings, dimensions and ordering codes, for selection and design work.
Questions answered in these manuals
What is the standard measurement accuracy of the ProcessMaster FEP300?
The standard accuracy is 0.4% of the measured value, with 0.2% available as an option.
From: ABB ProcessMaster FEP300 Electromagnetic Flowmeter Technical Specification
Which communication protocols does the ProcessMaster FEP300 support?
HART is standard; PROFIBUS PA and FOUNDATION Fieldbus are available as options.
From: ABB ProcessMaster FEP300 Electromagnetic Flowmeter Technical Specification
What is the maximum measuring error for the FEP300?
The device achieves a maximum measuring error of 0.2 % of rate using advanced filtering methods.
From: ABB ProcessMaster FEP300 Electromagnetic Flowmeter Data Sheet
Does the FEP300 require reconfiguration when changing transmitters?
No, revolutionary data storage enables transmitter interchange without the need for reconfiguration.
From: ABB ProcessMaster FEP300 Electromagnetic Flowmeter Data Sheet