Analog Devices ADAS1000/ADAS1000-1/ADAS1000-2 ECG Analog Front-End Datasheet

Analog Devices ADAS1000/ADAS1000-1/ADAS1000-2 ECG Analog Front-End Datasheet

Datasheet 80 pages · 1.8 MB · ZH

The Analog Devices ADAS1000, ADAS1000-1, and ADAS1000-2 are low-power ECG analog front-ends that measure ECG signals, thoracic impedance, pacemaker information, and lead connection or off status, then output the information as data frames at programmable data rates. They are suited to battery-powered portable applications and high-end diagnostic equipment.

ADAS1000 is a full-featured 5-channel ECG with respiration and pulse detection; ADAS1000-1 provides ECG channels without respiration and pulse detection, while ADAS1000-2 operates as a slave device. The devices are offered in 56-lead LFCSP and 64-lead LQFP packages and are rated from −40°C to +85°C. This manual is written in Chinese.

Read online

Analog Devices ADAS1000 / ADAS1000-1 / ADAS1000-2 specifications

Power consumption11 mW (1 lead), 15 mW (3 leads), 21 mW (all electrodes)
Front-end gain settingsGAIN 0 = ×1.4, GAIN 1 = ×2.1, GAIN 2 = ×2.8, GAIN 3 = ×4.2
CMRR at 10 Hz105 to 110 dB
Resolution19 bits (electrode/vector mode, 2 kHz data rate)
Package options56-lead LFCSP (9 mm × 9 mm), 64-lead LQFP (10 mm × 10 mm)
Operating temperature range−40°C to +85°C

Source: Analog Devices ADAS1000/ADAS1000-1/ADAS1000-2 ECG Analog Front-End Datasheet

All Analog Devices ADAS1000 / ADAS1000-1 / ADAS1000-2 documents by type

Datasheet 2

Ratings, dimensions and ordering codes, for selection and design work.

Questions answered in these manuals

Does ADAS1000-2 support respiration and pacemaker detection?

No. ADAS1000-2 operates as a slave device and provides ECG channels but has no respiration, pacemaker, or right-leg drive functions.

From: Analog Devices ADAS1000/ADAS1000-1/ADAS1000-2 ECG Analog Front-End Datasheet

What packages are available for the ADAS1000 series?

The devices are offered in 56-lead LFCSP and 64-lead LQFP packages.

From: Analog Devices ADAS1000/ADAS1000-1/ADAS1000-2 ECG Analog Front-End Datasheet

What standards does the ADAS1000 support?

It supports AAMI EC11:1991/(R)2001/(R)2007, EC13:2002/(R)2007, IEC60601-1 ed. 3.0b, 2005, IEC60601-2-25 ed. 1.0b, IEC60601-2-27 ed. 2.0, 2005 and IEC60601-2-51 ed. 1.0, 2005.

From: Analog Devices ADAS1000/ADAS1000-1/ADAS1000-2 Datasheet

What are the main applications for the ADAS1000?

Applications include Bedside Patient Monitoring, Portable Telemetry, Holter, AED, Cardiac Defibrillators, Ambulatory Monitors, and Pace Maker Programmer.

From: Analog Devices ADAS1000/ADAS1000-1/ADAS1000-2 Datasheet