Agilent 7890A-5975C and 1200SL-6410 Melamine Cyanuric Acid Analysis Application Note

This application note describes methods for analyzing the levels of melamine and cyanuric acid in pet food and related raw materials using the Agilent 7890A-5975C GC/MS and 1200SL-6410 Triple Quadrupole LC/MS. The GC/MS method employs extraction and then trimethylsilyl derivatization and is used here for screening. The detection limit is 10 µg/g. The LC/MS/MS method requires only extraction and then a simple isocratic elution on a normal-phase silica column (Rx-Sil) with water and acetonitrile mobile phase containing 5 mM ammonium acetate. The separation of the two compounds takes 5 min. The linear range is from 50 pg/mL to 50 ng/mL for melamine and from 1 ng/mL to 100 ng/mL for cyanuric acid. The LC/MS/MS method is demonstrated by analysis of the target compounds in complex food matrix.


Agilent 7890A-5975C and 1200SL-6410 Melamine Cyanuric Acid Analysis Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
185 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
D3D152B5DC3A7E899052246BC89CE6EC

About this manual

This application note describes a total solution for the analysis of melamine and cyanuric acid in pet food and related raw materials. It presents two methods: a screening method using GC/MS with trimethylsilyl derivatization and a confirmation/quantitation method using aqueous normal-phase LC/MS/MS without derivatization. The LC/MS/MS method offers higher sensitivity and selectivity, covering trace-level analysis in animal fluids to high levels in food matrices.

The document provides detailed experimental procedures, including sample preparation protocols for both GC/MS and LC/MS/MS. It specifies instrument conditions, column types, mobile phases, and mass spectrometry parameters (SIM for GC/MS and MRM for LC/MS/MS). Results demonstrate the linearity, detection limits, and performance of the methods on various food matrices such as pet food, wheat gluten, and corn meal.

What's inside

  • Abstract
  • Introduction
  • Sample Preparation
  • GC/MS Method Details
  • LC/MS Method Details
  • Results and Discussion
  • Selection of Monitoring Ions
  • Retention on Normal-Phase Column
  • Quantitative Analysis
  • Linearity and Detection Limits

Specifications

GC/MS Detection Limit10 ug/g
LC/MS/MS Linear Range Melamine50 pg/mL to 50 ng/mL
LC/MS/MS Linear Range Cyanuric Acid1 ng/mL to 100 ng/mL
LC/MS/MS Separation Time5 min
GC/MS ColumnDB-5MS 30 m x 0.25 mm x 0.25 ym
LC/MS/MS ColumnZORBAX Rx-Sil, 2.1 mm x 150 mm, 5 ym
LC/MS/MS Injection Volume10 pL
LC/MS/MS Source Temperature350°C

Frequently asked questions

What is the detection limit for the GC/MS screening method?

The detection limit for the GC/MS screening method is 10 ug/g.

Does the LC/MS/MS method require derivatization?

No, the LC/MS/MS method requires only extraction and then a simple isocratic elution without derivatization.

What mobile phase is used for the LC/MS/MS method?

The method uses water and acetonitrile mobile phase containing 5 mM ammonium acetate.

What software is used for the LC/MS/MS analysis?

The analysis uses MassHunter B.01.03 software.

Download PDF (185 KB)

More Agilent Manuals