Key facts
- Brand
- Agilent
- Model
- Turbo-V 81-M
- Category
- Other products
- Documents
- 2
- Languages
- EN
- Last added
- 2023-11-24
Start here

Agilent Turbo-V 81-M Turbomolecular Pump User Manual
Agilent Turbo-V 81-M is a turbomolecular pump for high and ultra-high vacuum applications. It can pump any type of non-corrosive gas or gas compound, is not suitable for pumping liquids or solid particles, and uses a high speed turbine with a maximum speed of 80000 rpm. It is free of contaminating agents and is suitable for applications requiring a clean vacuum.
The manual covers safety guidelines, storage, preparation for installation, installation, use, maintenance, disposal, and technical information including accessories and spare parts. It states operating and storage environmental limits, heating limits, venting requirements, and controller compatibility.
Agilent Turbo-V 81-M specifications
| Maximum turbine speed | 80000 rpm |
|---|---|
| Storage temperature range | -20 °C to 70 °C |
| Storage relative humidity range | 0 to 95 % (non condensing) |
| Shelf life | 10 months from shipping date |
| Operating temperature | +5 °C to +35 °C |
| Operating relative humidity | 0 – 95 % (non-condensing) |
| Maximum pressure | 2 bar above atmospheric pressure |
| Maximum inlet flange temperature (ConFlat) | 120 °C |
| Maximum inlet flange temperature (ISO) | 80 °C |
| Vent port pressure | less than 2 bar above atmospheric pressure |
| Counter-flange torque capacity | 250 Nm |
| ISO 63 flange fixing | 4 M10 clamps, 22 Nm |
Source: Agilent Turbo-V 81-M Turbomolecular Pump User Manual
All Agilent Turbo-V 81-M documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
What gases can the Turbo-V 81-M pump?
It can pump any type of non-corrosive gas or gas compound, and it is not suitable for pumping liquids or solid particles.
Does the Turbo-V 81-M require maintenance?
The Turbo-V 81-M series pumps does not require any maintenance.
What are the applications for the Turbo-V 81-M?
The pump is suitable for mass spectrometers, electron microscopes, surface analysis, leak detection, and research laboratories.
Does the Turbo-V 81-M require lubrication?
No, it uses Varian's unique dry lubricant technology combined with ceramic bearings, making it a maintenance-free pump.